ULTRASOUND TRANSMISSION MEASUREMENTS THROUGH THE OS CALCIS

被引:119
|
作者
ZAGZEBSKI, JA [1 ]
ROSSMAN, PJ [1 ]
MESINA, C [1 ]
MAZESS, RB [1 ]
MADSEN, EL [1 ]
机构
[1] UNIV WISCONSIN, DEPT MED PHYS, MADISON, WI 53706 USA
关键词
ULTRASOUND ATTENUATION; SPEED-OF-SOUND; BONE; OSTEOPOROSIS;
D O I
10.1007/BF02565130
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A method of measuring ultrasonic propagation in the os calcis was devised for assessing bone properties in humans. Speed-of-sound (SOS) and broadband ultrasound attenuation (BUA) were measured using broadband acoustic pulses transmitted and received by a pair of focused transducers. The transducers are mounted coaxially in a water tank with the subject's heel in between. Reproducibility of results in an adult male was 10% for the BUA and 1.2% for the SOS. Both SOS and BUA changed when the transmission path through the os calcis was varied. For a population of normal male subjects, SOS and BUA were correlated with densitometry results on the os calcis, but less well correlated to area density at remote sites.