共 50 条
ULTRASOUND TRANSMISSION MEASUREMENTS THROUGH THE OS CALCIS
被引:119
|作者:
ZAGZEBSKI, JA
[1
]
ROSSMAN, PJ
[1
]
MESINA, C
[1
]
MAZESS, RB
[1
]
MADSEN, EL
[1
]
机构:
[1] UNIV WISCONSIN, DEPT MED PHYS, MADISON, WI 53706 USA
关键词:
ULTRASOUND ATTENUATION;
SPEED-OF-SOUND;
BONE;
OSTEOPOROSIS;
D O I:
10.1007/BF02565130
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
A method of measuring ultrasonic propagation in the os calcis was devised for assessing bone properties in humans. Speed-of-sound (SOS) and broadband ultrasound attenuation (BUA) were measured using broadband acoustic pulses transmitted and received by a pair of focused transducers. The transducers are mounted coaxially in a water tank with the subject's heel in between. Reproducibility of results in an adult male was 10% for the BUA and 1.2% for the SOS. Both SOS and BUA changed when the transmission path through the os calcis was varied. For a population of normal male subjects, SOS and BUA were correlated with densitometry results on the os calcis, but less well correlated to area density at remote sites.
引用
收藏
页码:107 / 111
页数:5
相关论文