FINGERS - 3-DIMENSIONAL MR-IMAGING AND ANGIOGRAPHY WITH A LOCAL GRADIENT COIL

被引:14
|
作者
BLACKBAND, SJ [1 ]
CHAKRABARTI, I [1 ]
GIBBS, P [1 ]
BUCKLEY, DL [1 ]
HORSMAN, A [1 ]
机构
[1] UNIV NEWCASTLE UPON TYNE,DEPT ORTHOPAED,NEWCASTLE TYNE,ENGLAND
关键词
ARTERIES; EXTREMITIES; FINGERS AND TOES; HAND; MR;
D O I
10.1148/radiology.190.3.8115647
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors interfaced a 12-cm-diameter local gradient coil with a solenoid radio-frequency coil (2.2-cm internal diameter, 1.4-cm long) to a recalibrated clinical magnetic resonance imaging system. Three-dimensional gradient-echo imaging and phase-contrast angiography were performed on the fingers of five healthy volunteers (four men and one woman, aged 23-33 years). Data sets (256 x 256 x 28) were acquired with spatial resolution of 100 x 100 x 500 mu m in 6 minutes 51 seconds. The angiograms clearly delineated blood vessels that were less than 1 mm in diameter.
引用
收藏
页码:895 / 899
页数:5
相关论文
共 50 条
  • [1] 3-DIMENSIONAL TIME-OF-FLIGHT MR ANGIOGRAPHY WITH A SPECIALIZED GRADIENT HEAD COIL
    TKACH, JA
    RUGGIERI, PM
    DILLINGER, JJ
    ROSS, JS
    MODIC, MT
    MASARYK, TJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (02): : 365 - 375
  • [2] CAROTID AND VERTEBRAL ARTERY DISSECTIONS - 3-DIMENSIONAL TIME-OF-FLIGHT MR-ANGIOGRAPHY AND MR-IMAGING VERSUS CONVENTIONAL ANGIOGRAPHY
    LEVY, C
    LAISSY, JP
    RAVEAU, V
    AMARENCO, P
    SERVOIS, V
    BOUSSER, MG
    TUBIANA, JM
    [J]. RADIOLOGY, 1994, 190 (01) : 97 - 103
  • [3] RAPID 3-DIMENSIONAL ANGIOGRAPHY WITH UNDERSAMPLED MR IMAGING
    WEDEEN, VJ
    CHAO, YS
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (01) : 24 - 30
  • [4] ERRORS IN QUANTITATIVE DYNAMIC 3-DIMENSIONAL KEYHOLE MR-IMAGING OF THE BREAST
    PLEWES, DB
    BISHOP, J
    SOUTAR, I
    COHEN, E
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (03): : 361 - 364
  • [5] HIGH-RESOLUTION, SHORT ECHO TIME MR IMAGING OF THE FINGERS AND WRIST WITH A LOCAL GRADIENT COIL
    WONG, EC
    JESMANOWICZ, A
    HYDE, JS
    [J]. RADIOLOGY, 1991, 181 (02) : 393 - 397
  • [6] FLASE - 3-DIMENSIONAL, SHORT TR, SE MR-IMAGING PULSE SEQUENCE
    MA, J
    WEHRLI, FW
    [J]. RADIOLOGY, 1995, 197 : 389 - 389
  • [7] 3-DIMENSIONAL ANISOTROPY CONTRAST MR-IMAGING OF THE RAT NERVOUS-SYSTEM - MR HISTOLOGY
    NAKADA, T
    MATSUZAWA, H
    [J]. RADIOLOGY, 1995, 194 (02) : 628 - 628
  • [8] MR IMAGING OF THE AORTA WITH 3-DIMENSIONAL VESSEL RECONSTRUCTION - VALIDATION BY ANGIOGRAPHY
    VALK, PE
    HALE, JD
    KAUFMAN, L
    CROOKS, LE
    HIGGINS, CB
    [J]. RADIOLOGY, 1985, 157 (03) : 721 - 725
  • [9] MR-IMAGING OF THE KNEE IN THE SAGITTAL PROJECTION - COMPARISON OF 3-DIMENSIONAL GRADIENT-ECHO AND SPIN-ECHO SEQUENCES
    REEDER, JD
    MATZ, SO
    BECKER, L
    ANDELMAN, SM
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 153 (03) : 537 - 540
  • [10] DETECTION OF KNEE HYALINE CARTILAGE DEFECTS USING FAT-SUPPRESSED 3-DIMENSIONAL SPOILED GRADIENT-ECHO MR-IMAGING - COMPARISON WITH STANDARD MR-IMAGING AND CORRELATION WITH ARTHROSCOPY
    DISLER, DG
    MCCAULEY, TR
    WIRTH, CR
    FUCHS, MD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (02) : 377 - 382