LABORATORY AND CLINICAL-EVALUATION OF WHITE BLOOD-CELL DIFFERENTIAL COUNTS - COMPARISON OF THE COULTER VCS, TECHNICON H-1, AND 800-CELL MANUAL METHOD

被引:44
|
作者
SWAIM, WR
机构
[1] UNIV MINNESOTA,SCH MED,DEPT LAB MED & PATHOL,MINNEAPOLIS,MN 55455
[2] UNIV MINNESOTA,SCH MED,DEPT MED,MINNEAPOLIS,MN 55455
关键词
COULTER VCS; TECHNICON H-1; DIFFERENTIAL LEUKOCYTE COUNT; DEMING REGRESSION;
D O I
10.1093/ajcp/95.3.381
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Eight hundred-cell manual white blood cell differential count evaluations of Coulter VCSTM and Technicon H-1TM included estimates of accuracy, imprecision, random and systematic errors, clinical relevance, and detection of immature neutrophils. Accuracy was acceptable, except for H-1 testing of monocytes. Instrument imprecision was similar, except for tighter monocyte values as measured by the VCS. Deming regression determined random errors were greater for H-1 monocytes and proportional errors were similar. Constant errors for the VCS were less than the H-1 for neutrophils and monocytes and greater than the H-1 for eosinophils and lymphocytes. H-1 morphologic false-negative rates were twice those for VCS. True-positive and false-negative rates in cases with immature neutrophils were 70.6% and 20.3% for the VCS and 35% and 55.9% for the H-1. The reference false-positive rate was 4.4%. Clinically appropriate VCS flags were generated in distributionally abnormal cases. Predictive values were higher for the VCS.