A comparison study between the Siemens ADVIA 120 and manual method for the differential white blood cell count in goats

被引:0
|
作者
Oikonomidis, Ioannis L. [1 ,2 ]
Brozos, Christos [3 ]
Tsouloufi, Theodora K. [2 ]
Kiossis, Evangelos [3 ]
Kritsepi-Konstantinou, Maria [2 ]
机构
[1] Univ Liverpool, Inst Infect Vet & Ecol Sci, Dept Vet Anat Physiol & Pathol, Liverpool, England
[2] Aristotle Univ Thessaloniki, Fac Hlth Sci, Sch Vet Med, Diagnost Lab, Thessaloniki, Greece
[3] Aristotle Univ Thessaloniki, Fac Hlth Sci, Farm Anim Clin, Sch Vet Med, Thessaloniki, Greece
关键词
blood smear; caprine; complete blood count; hematology; leukocyte; leukogram; CATS;
D O I
10.1111/vcp.13337
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: Although widely used, the ADVIA 120 hematology analyzer has not been previously validated for determining the differential leukocyte count in goats. Objectives: The aim of this study was to compare the differential leukocyte counts provided by the ADVIA 120 (A-diff) and the manual method (M-Diff) in goats. Methods: EDTA blood samples that were analyzed within 4 h of collection were used in the study. The following exclusion criteria were applied: inappropriately filled tubes or tubes containing clots, erroneous ADVIA peroxidase cytograms, and blood smears of poor quality. The A-Diff was compared with the M-Diff performed by two independent observers on 200 leukocytes. Results: Forty samples were included after previously excluding eight samples. The correlation between the A-Diff and M-Diff was very strong for eosinophils (r = .870, p < .001) and strong for lymphocytes (r = .796, p < .001) and neutrophils (r = .730, p < .001), while no significant correlation was observed for monocytes (r = .026, p = .872). The Passing-Bablok regression analyses revealed statistically significant constant errors for neutrophils (5.83%; 95% confidence interval [CI]: 0.41%, 12.18%) and eosinophils (1.89%; 95% CI: 1.17%, 2.71%). Bland-Altman analyses showed a statistically significant negative bias for lymphocytes (-5.0%) and a statistically significant positive bias for eosinophils (2.2%). The very low basophil percentages precluded a meaningful method comparison. Conclusions: The ADVIA 120 overall demonstrated good performance for the differential WBC count in goats under the conditions of this study. Therefore, it can be considered suitable for routine hematologic screening in goats. Nonetheless, it should be emphasized that any abnormal result should be confirmed with a blood smear evaluation.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 50 条
  • [1] A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep
    Oikonomidis, Ioannis L.
    Brozos, Christos
    Kiossis, Evangelos
    Tsouloufi, Theodora K.
    Kritsepi-Konstantinou, Maria
    [J]. VETERINARY CLINICAL PATHOLOGY, 2021, 50 (02) : 203 - 208
  • [2] Differential white blood cell counts in rabbits: a comparison of the Advia 2120 and a manual method
    Oikonomidis, Ioannis L.
    Milne, Elspeth
    Piccinelli, Chiara
    [J]. JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2021, 33 (04) : 670 - 676
  • [3] Nucleated red blood cell count and blast count provided by the new software of the Siemens ADVIA® 2120:: comparison with the microscope reference method
    Zini, G.
    Valentini, S.
    De Padua, L.
    Vannata, B.
    Di Mario, A.
    Garzia, M.
    d'Onofrio, G.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 : 81 - 81
  • [4] Measurement uncertainty in manual white blood cell differential count
    Vilcnik, T. Zuran
    [J]. CLINICA CHIMICA ACTA, 2024, 558
  • [5] Evaluation of equine hemograms using the ADVIA 120 as compared with an impedance counter and manual differential count
    Giordano, Alessia
    Rossi, Gabriele
    Pieralisi, Cristina
    Paltrinieri, Saverio
    [J]. VETERINARY CLINICAL PATHOLOGY, 2008, 37 (01) : 21 - 30
  • [6] WHITE BLOOD-CELL DIFFERENTIAL COUNT - COMPARISON OF MANUAL SPREAD FILMS AND SEMIAUTOMATED SPUN FILMS
    COYNE, RL
    [J]. CANADIAN JOURNAL OF MEDICAL TECHNOLOGY, 1976, 38 (01): : 18 - 25
  • [7] Canine complete blood counts: a comparison of four in-office instruments with the ADVIA 120 and manual differential counts
    Welles, Elizabeth G.
    Hall, Amy S.
    Carpenter, D. Mark
    [J]. VETERINARY CLINICAL PATHOLOGY, 2009, 38 (01) : 20 - 29
  • [8] Estimation of white blood cell and platelet counts in ovine blood smears, and a comparison with the ADVIA 120 hematology analyzer
    Katsogiannou, Eleni G.
    Athanasiou, Labrini V.
    Katsoulos, Panagiotis D.
    Polizopoulou, Zoe S.
    Tzivara, Athanasia
    Christodoulopoulos, Georgios
    [J]. VETERINARY CLINICAL PATHOLOGY, 2020, 49 (02) : 222 - 226
  • [9] Relationship between rotating shift work and white blood cell count, white blood cell differential count, obesity, and metabolic syndrome of nurses
    Chang, Wen-Pei
    Lin, Yen-Kuang
    [J]. CHRONOBIOLOGY INTERNATIONAL, 2022, 39 (02) : 159 - 168
  • [10] Accurate white blood cell differential by Alinity hq: A comparison with flow cytometry and manual differential
    Gambell, Peter
    Rowley, Grant
    Pham, Tracy Anh Thu
    Dang, Theresa Loan
    Mulumba, Huguette
    Smith, Larry
    Lakos, Gabriella
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (02) : 288 - 295