A comparison study between the Siemens ADVIA 120 and manual method for the differential white blood cell count in goats

被引:0
|
作者
Oikonomidis, Ioannis L. [1 ,2 ]
Brozos, Christos [3 ]
Tsouloufi, Theodora K. [2 ]
Kiossis, Evangelos [3 ]
Kritsepi-Konstantinou, Maria [2 ]
机构
[1] Univ Liverpool, Inst Infect Vet & Ecol Sci, Dept Vet Anat Physiol & Pathol, Liverpool, England
[2] Aristotle Univ Thessaloniki, Fac Hlth Sci, Sch Vet Med, Diagnost Lab, Thessaloniki, Greece
[3] Aristotle Univ Thessaloniki, Fac Hlth Sci, Farm Anim Clin, Sch Vet Med, Thessaloniki, Greece
关键词
blood smear; caprine; complete blood count; hematology; leukocyte; leukogram; CATS;
D O I
10.1111/vcp.13337
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: Although widely used, the ADVIA 120 hematology analyzer has not been previously validated for determining the differential leukocyte count in goats. Objectives: The aim of this study was to compare the differential leukocyte counts provided by the ADVIA 120 (A-diff) and the manual method (M-Diff) in goats. Methods: EDTA blood samples that were analyzed within 4 h of collection were used in the study. The following exclusion criteria were applied: inappropriately filled tubes or tubes containing clots, erroneous ADVIA peroxidase cytograms, and blood smears of poor quality. The A-Diff was compared with the M-Diff performed by two independent observers on 200 leukocytes. Results: Forty samples were included after previously excluding eight samples. The correlation between the A-Diff and M-Diff was very strong for eosinophils (r = .870, p < .001) and strong for lymphocytes (r = .796, p < .001) and neutrophils (r = .730, p < .001), while no significant correlation was observed for monocytes (r = .026, p = .872). The Passing-Bablok regression analyses revealed statistically significant constant errors for neutrophils (5.83%; 95% confidence interval [CI]: 0.41%, 12.18%) and eosinophils (1.89%; 95% CI: 1.17%, 2.71%). Bland-Altman analyses showed a statistically significant negative bias for lymphocytes (-5.0%) and a statistically significant positive bias for eosinophils (2.2%). The very low basophil percentages precluded a meaningful method comparison. Conclusions: The ADVIA 120 overall demonstrated good performance for the differential WBC count in goats under the conditions of this study. Therefore, it can be considered suitable for routine hematologic screening in goats. Nonetheless, it should be emphasized that any abnormal result should be confirmed with a blood smear evaluation.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 50 条
  • [31] A link between white blood cell count and blood pressure levels
    Kazushi Tsuda
    [J]. Hypertension Research, 2024, 47 : 537 - 539
  • [32] A Comparison between a Manual and Automatic Method to Characterize Red Blood Cell Trajectories
    Pinho, Diana
    Gayubo, Fernando
    Pereira, Ana I.
    Lima, Rui
    [J]. NUMERICAL ANALYSIS AND APPLIED MATHEMATICS ICNAAM 2011: INTERNATIONAL CONFERENCE ON NUMERICAL ANALYSIS AND APPLIED MATHEMATICS, VOLS A-C, 2011, 1389
  • [33] A link between white blood cell count and blood pressure levels
    Tsuda, Kazushi
    [J]. HYPERTENSION RESEARCH, 2024, 47 (02) : 537 - 539
  • [34] Comparison of the anticoagulants K2-EDTA and K3-EDTA for red blood cell count using ADVIA 120 hematology system
    Krnjak, L.
    Lenart, K.
    Jerin, A.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 : 70 - 71
  • [35] CONCERNING AUTOMATION OF THE WHITE BLOOD-CELL DIFFERENTIAL COUNT - REPLY
    PRESTON, K
    [J]. COMPUTER, 1983, 16 (08) : 9 - 10
  • [36] Comparison of manual microscope method and an automated cell count in serous fluids
    Angeleri, Anabela
    Ariagno, Julia
    Sardi, Melba
    Carbi, Claudio
    Palaoro, Luis
    Rocher, Adriana
    [J]. ACTA BIOQUIMICA CLINICA LATINOAMERICANA, 2017, 51 (01): : 37 - 44
  • [37] DIFFERENTIAL WHITE BLOOD CELL COUNT AND METABOLIC SYNDROME IN JAPANESE MEN
    Ikeda, Y.
    Hisakawa, N.
    Ohguro, T.
    Nishiuchi, J.
    Kumon, Y.
    [J]. ATHEROSCLEROSIS, 2016, 252 : E181 - E182
  • [38] COMPARISON BETWEEN MANUAL AND AUTOMATED CELL COUNT METHODS IN THE DIAGNOSIS OF SBP
    Waddah, Mohamed
    Bendall, Oliver
    Moodley, Prebashan
    Bennett, Kris
    Chan, Sally
    Keelty, Nigel
    Steer, Jane
    Siewruk, John
    Thomas, Wayne
    Tilley, Robert
    Haro, Juan Acevedo
    [J]. GUT, 2023, 72 : A59 - A59
  • [39] COMPARISON BETWEEN MANUAL AND AUTOMATED CELL COUNT METHODS IN THE DIAGNOSIS OF SBP
    Waddah, Mohamed
    Bendall, Oliver
    Moodley, Prebashan
    Bennett, Kris
    Chan, Sally
    Keelty, Nigel
    Steer, Jane
    Siewruk, John
    Thomas, Wayne
    Tilley, Robert
    Haro, Juan Acevedo
    [J]. GUT, 2023, 72 (SUPPL_3) : A59 - A59
  • [40] Relationship between white blood cell count and incident hypertension
    Shankar, A
    Klein, BEK
    Klein, R
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2004, 17 (03) : 233 - 239