WHICH MONITOR - AN ANALYSIS OF 2000 INCIDENT REPORTS

被引:127
|
作者
WEBB, RK
VANDERWALT, JH
RUNCIMAN, WB
WILLIAMSON, JA
COCKINGS, J
RUSSELL, WJ
HELPS, S
机构
[1] ROYAL ADELAIDE HOSP,ADELAIDE,SA 5000,AUSTRALIA
[2] UNIV ADELAIDE,DEPT ANAESTHESIA & INTENS CARE,HYPERBAR MED UNIT,ADELAIDE,SA 5001,AUSTRALIA
[3] WOMENS & CHILDRENS HOSP,DEPT PAEDIAT ANAESTHESIA,ADELAIDE,SA,AUSTRALIA
关键词
ANESTHESIA; EQUIPMENT; PULSE OXIMETER; CAPNOGRAPH; OXYGEN SUPPLY FAILURE ALARM; OXYGEN ANALYZER; ELECTROCARDIOGRAPH; LOW PRESSURE ALARM MONITOR; PATIENT TEMPERATURE; NERVE STIMULATOR; STETHOSCOPE;
D O I
10.1177/0310057X9302100508
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The role of monitors in patients undergoing general anaesthesia was studied by analysing the first 2000 incidents reported to the Australian Incident Monitoring Study; 1256 (63%) were considered applicable to this study. In 52% of these a monitor detected the incident first; oximetry (27%) and capnography (24%) detected over half of the monitor detected incidents, the electrocardiograph 19%, bloodpressure monitors 12%, a low pressure (circuit) alarm 8%, and the oxygen analyser 4%. Of the other monitors used, 5 first detected 1-2% of incidents, and the remaining 8 less than 0.5% each. The oximeter would have detected over 40% of the monitor detected incidents had its more informative modulated pulse tone always been relied upon instead of the ''bleep'' of the ECG. A theoretical analysis was then carried out to determine which of an array of 17 monitors would reliably have detected each incident had each monitor been used on its own and had the incident been allowed to evolve. To facilitate ''scoring'' of monitors, the incidents were categorized empirically into 60 clinical situations; 40% of applicable incidents were ''counted for by only 5 clinical situations, 60% by 10 and nearly 80% by 20. 98% were accounted for by the 60 situations. A pulse oximeter, used on its own, would theoretically have detected 82% of applicable incidents (nearly 60% before any potential for organ damage). These figures for capnography are 55% and 43% and for oximetry and capnography combined are 88% and 65%, respectively. With the addition of blood pressure monitoring these become 93% and 65%, and of an oxygen analyser, 95 and 67%. Other monitors, including the ECG, each increase the yield by by less than 0.5%. The international monitoring recommendations and those of the Australian and New Zealand College of Anaesthetists are thoroughly vindicated by the patterns revealed in this study. The priority sequence of monitor acquisition for those with limited resources should be stethoscope, sphygmomanometer, oxygen analyser if nitrous oxide is to be used, pulse oximeter capnograph, high pressure alarm, and, if patients are to be mechanically ventilated, a low pressure alarm (or spirometer with alarm); an ECG, a defibrillator, a spirometer and a thermometer should be available.
引用
下载
收藏
页码:529 / 542
页数:14
相关论文
共 50 条
  • [31] Perianesthetic dental injuries: Analysis of incident reports
    Givol, N
    Gershtansky, Y
    Halamish-Shani, T
    Taicher, S
    Perel, A
    Segal, E
    JOURNAL OF CLINICAL ANESTHESIA, 2004, 16 (03) : 173 - 176
  • [32] Endobronchial intubation: An analysis of 4000 incident reports
    McCoy, EP
    Russell, WJ
    ANESTHESIOLOGY, 1996, 85 (3A) : A914 - A914
  • [33] INCIDENT REPORTS
    DOLLE, R
    JOURNAL OF ENVIRONMENTAL HEALTH, 1993, 56 (05) : 43 - 44
  • [34] THE 'MONITOR REPORTS'
    MAYO, EL
    NEW LETTERS, 1981, 47 (2-3): : 148 - 148
  • [35] Which Monitor?
    M P Samuels
    Pediatric Research, 1999, 45 : 13 - 13
  • [36] Which monitor?
    Samuels, MP
    PEDIATRIC RESEARCH, 1999, 45 (05) : 13A - 13A
  • [37] Safety incident reporting in emergency radiology: analysis of 1717 safety incident reports
    Mansouri M.
    Shaqdan K.W.
    Aran S.
    Raja A.S.
    Lev M.H.
    Abujudeh H.H.
    Emergency Radiology, 2015, 22 (6) : 623 - 630
  • [38] Outpatient Safety in the Elderly Population: An Analysis of Incident Reports
    Yang, J. Y.
    Del Rosario, J.
    Sharma, A.
    JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 2020, 68 : S330 - S331
  • [39] Characteristics of events in which police responded to overdoses: an examination of incident reports in Rhode Island
    Alexandria Macmadu
    Annajane Yolken
    Lisa Frueh
    Jai’el R. Toussaint
    Roxxanne Newman
    Brendan P. Jacka
    Alexandra B. Collins
    Brandon D. L. Marshall
    Harm Reduction Journal, 19
  • [40] Analysis of incident reports in an anesthesiology unit of a university hospital
    Marfan, Luis
    Pedemonte, Juan C.
    Sandoval, Daniela
    Ferdinand, Constanza
    Camus, Lorena
    Lacassie, Hector J.
    REVISTA MEDICA DE CHILE, 2017, 145 (04) : 441 - 448