WHICH MONITOR - AN ANALYSIS OF 2000 INCIDENT REPORTS

被引:127
|
作者
WEBB, RK
VANDERWALT, JH
RUNCIMAN, WB
WILLIAMSON, JA
COCKINGS, J
RUSSELL, WJ
HELPS, S
机构
[1] ROYAL ADELAIDE HOSP,ADELAIDE,SA 5000,AUSTRALIA
[2] UNIV ADELAIDE,DEPT ANAESTHESIA & INTENS CARE,HYPERBAR MED UNIT,ADELAIDE,SA 5001,AUSTRALIA
[3] WOMENS & CHILDRENS HOSP,DEPT PAEDIAT ANAESTHESIA,ADELAIDE,SA,AUSTRALIA
关键词
ANESTHESIA; EQUIPMENT; PULSE OXIMETER; CAPNOGRAPH; OXYGEN SUPPLY FAILURE ALARM; OXYGEN ANALYZER; ELECTROCARDIOGRAPH; LOW PRESSURE ALARM MONITOR; PATIENT TEMPERATURE; NERVE STIMULATOR; STETHOSCOPE;
D O I
10.1177/0310057X9302100508
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The role of monitors in patients undergoing general anaesthesia was studied by analysing the first 2000 incidents reported to the Australian Incident Monitoring Study; 1256 (63%) were considered applicable to this study. In 52% of these a monitor detected the incident first; oximetry (27%) and capnography (24%) detected over half of the monitor detected incidents, the electrocardiograph 19%, bloodpressure monitors 12%, a low pressure (circuit) alarm 8%, and the oxygen analyser 4%. Of the other monitors used, 5 first detected 1-2% of incidents, and the remaining 8 less than 0.5% each. The oximeter would have detected over 40% of the monitor detected incidents had its more informative modulated pulse tone always been relied upon instead of the ''bleep'' of the ECG. A theoretical analysis was then carried out to determine which of an array of 17 monitors would reliably have detected each incident had each monitor been used on its own and had the incident been allowed to evolve. To facilitate ''scoring'' of monitors, the incidents were categorized empirically into 60 clinical situations; 40% of applicable incidents were ''counted for by only 5 clinical situations, 60% by 10 and nearly 80% by 20. 98% were accounted for by the 60 situations. A pulse oximeter, used on its own, would theoretically have detected 82% of applicable incidents (nearly 60% before any potential for organ damage). These figures for capnography are 55% and 43% and for oximetry and capnography combined are 88% and 65%, respectively. With the addition of blood pressure monitoring these become 93% and 65%, and of an oxygen analyser, 95 and 67%. Other monitors, including the ECG, each increase the yield by by less than 0.5%. The international monitoring recommendations and those of the Australian and New Zealand College of Anaesthetists are thoroughly vindicated by the patterns revealed in this study. The priority sequence of monitor acquisition for those with limited resources should be stethoscope, sphygmomanometer, oxygen analyser if nitrous oxide is to be used, pulse oximeter capnograph, high pressure alarm, and, if patients are to be mechanically ventilated, a low pressure alarm (or spirometer with alarm); an ECG, a defibrillator, a spirometer and a thermometer should be available.
引用
收藏
页码:529 / 542
页数:14
相关论文
共 50 条
  • [21] PATIENT AWARENESS DURING ANESTHESIA - AN ANALYSIS OF 2000 INCIDENT REPORTS
    OSBORNE, GA
    WEBB, RK
    RUNCIMAN, WB
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 653 - 654
  • [22] THE OXYGEN ANALYZER - APPLICATIONS AND LIMITATIONS - AN ANALYSIS OF 2000 INCIDENT REPORTS
    BARKER, L
    WEBB, RK
    RUNCIMAN, WB
    VANDERWALT, JH
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 570 - 574
  • [23] PHYSICAL INJURIES AND ENVIRONMENTAL SAFETY IN ANESTHESIA - AN ANALYSIS OF 2000 INCIDENT REPORTS
    SINGLETON, RJ
    LUDBROOK, GL
    WEBB, RK
    FOX, MAL
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 659 - 663
  • [24] PROBLEMS RELATED TO THE ENDOTRACHEAL-TUBE - AN ANALYSIS OF 2000 INCIDENT REPORTS
    SZEKELY, SM
    WEBB, RK
    WILLIAMSON, JA
    RUSSELL, WJ
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 611 - 616
  • [25] THE WRONG DRUG-PROBLEM IN ANESTHESIA - AN ANALYSIS OF 2000 INCIDENT REPORTS
    CURRIE, M
    MACKAY, P
    MORGAN, C
    RUNCIMAN, WB
    RUSSELL, WJ
    SELLEN, A
    WEBB, RK
    WILLIAMSON, JA
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 596 - 601
  • [26] BLOOD-PRESSURE MONITORING - APPLICATIONS AND LIMITATIONS - AN ANALYSIS OF 2000 INCIDENT REPORTS
    COCKINGS, JGL
    WEBB, RK
    KLEPPER, ID
    CURRIE, M
    MORGAN, C
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 565 - 569
  • [27] THE CLASSIFICATION AND ANALYSIS OF RAILWAY INCIDENT REPORTS
    Gibson, W. Huw
    Mills, Ann
    Hesketh, Sarah
    RAIL HUMAN FACTORS AROUND THE WORLD: IMPACTS ON AND OF PEOPLE FOR SUCCESSFUL RAIL OPERATIONS, 2012, : 11 - 19
  • [28] RECOVERY ROOM INCIDENTS IN THE 1ST 2000 INCIDENT REPORTS
    VANDERWALT, JH
    WEBB, RK
    OSBORNE, GA
    MORGAN, C
    MACKAY, P
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) : 650 - 652
  • [29] Formal Concept Analysis of Medical Incident Reports
    Baba, Takahiro
    Liu, Lucing
    Hirokawa, Sachio
    KNOWLEDGE-BASED AND INTELLIGENT INFORMATION AND ENGINEERING SYSTEMS, PT III, 2010, 6278 : 207 - +
  • [30] Analysis of Incident Reports of a Dental University Hospital
    Kimura, Yasuyuki
    Tonami, Ken-ichi
    Toyofuku, Akira
    Nitta, Hiroshi
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (16)