Metal Artefact Reduction by Dual-energy Computed Tomography Using Monoenergetic Extrapolation: In-vitro Determination of Optimal Monoenergetic Level with Different Metallic Implants Using a Phantom Body

被引:2
|
作者
Chan, W. C. S. [1 ]
Tsang, J. P. K. [1 ]
Wong, W. Y. [1 ]
Chu, P. Y. [1 ]
To, V. Y. K. [1 ]
Lee, C. Y. [1 ]
Yeung, T. W. [1 ]
Leung, O. C. [1 ]
Sin, N. Y. [1 ]
Li, O. C. [1 ]
Yuen, M. K. [1 ]
机构
[1] Tuen Mun Hosp, Dept Radiol, Tuen Mun, Hong Kong, Peoples R China
来源
HONG KONG JOURNAL OF RADIOLOGY | 2016年 / 19卷 / 01期
关键词
Artifacts; Metals; Prostheses and implants; Tomography; X-ray computed;
D O I
10.12809/hkjr1615356
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To identify the optimal monoenergetic level, balancing metal artefacts, and the amount of noise present for imaging of metal implants using dual-energy computed tomography (CT) and focusing on the assessment of peri-prosthetic soft tissue. Methods: Four metallic implants commonly used in the hips were placed in a phantom body: unipolar hemiprosthesis, dynamic hip screw (DHS), intra-medullary (IM) nail, and titanium insert. The unipolar hemiprosthesis was imaged at two points: the head and stem. The head of the hemiprosthesis and DHS were imaged in two axes: one axial to and one resembling the angle at its expected position in the hip with respect to the scanner. The IM nail was assessed both at the level with and without a screw inserted. A region of interest to measure the noise level of the images was first performed with different monoenergetic levels (70-170 kV with increments of 10 kV). Four monoenergetic levels were then chosen (80, 90, 105, 120 kV) for each implant and were assessed and scored (presence of least to most artefacts: score 1-4) by nine radiologists who were blinded to the monoenergetic level. A total of eight sets of images were assessed. The scores for different monoenergetic levels were compared using analysis of variance. Results: In the first part of the experiment, the images with the least amount of noise were in the range of 85-95 kV, thus we included 90 kV among the images for subsequent scoring. The mean score for different monoenergetic levels for all implants was as follows: 3.94 for 80 kV, 2.68 for 90 kV, 1.50 for 105 kV, and 1.88 for 120 kV (p < 0.001), with 105 kV having the least metal artefacts. For subgroup analysis of individual implants, 105 kV was found to produce the best quality images with a statistically significant better score for hip stem, DHS, and IM nail. 120 kV trended towards being the best monoenergetic level when imaging the hip head and the IM nail with screw where relatively more artefacts were present. 90 kV trended towards being the best monoenergetic level when imaging the titanium insert where artefacts were nearly absent. Conclusion: With regard to imaging the soft tissue around a metallic implant, the overall optimal monoenergetic level for reduction of metal artefacts using dual-energy CT is 105 kV. When more artefacts are inherently present, 120 kV trended towards being the best monoenergetic level. When artefacts are minimal, 90 kV trended towards being the optimal monoenergetic level with the least amount of noise present.
引用
收藏
页码:35 / 42
页数:8
相关论文
共 50 条
  • [21] Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses
    Nicola Magarelli
    Vincenzo De Santis
    Giammaria Marziali
    Amerigo Menghi
    Aaron Burrofato
    Luigi Pedone
    Dario Del Prete
    Roberto Iezzi
    Chiara de Waure
    Marianna D’andrea
    Antonio Leone
    Cesare Colosimo
    La radiologia medica, 2018, 123 : 593 - 600
  • [22] Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses
    Magarelli, Nicola
    De Santis, Vincenzo
    Marziali, Giammaria
    Menghi, Amerigo
    Burrofato, Aaron
    Pedone, Luigi
    Del Prete, Dario
    Iezzi, Roberto
    de Waure, Chiara
    D'andrea, Marianna
    Leone, Antonio
    Colosimo, Cesare
    RADIOLOGIA MEDICA, 2018, 123 (08): : 593 - 600
  • [23] Dual layer computed tomography: Reduction of metal artefacts from posterior spinal fusion using virtual monoenergetic imaging
    Dangelmaier, Julia
    Schwaiger, Benedikt J.
    Gersing, Alexandra S.
    Kopp, Felix F.
    Sauter, Andreas
    Renz, Martin
    Riederer, Isabelle
    Braren, Rickmer
    Pfeiffer, Daniela
    Fingerle, Alexander
    Rummeny, Ernst J.
    Noel, Peter B.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 105 : 195 - 203
  • [24] Metal artifact reduction in cervix brachytherapy with titanium applicators using dual-energy CT through virtual monoenergetic images and an iterative algorithm: A phantom study
    Kanani, Abolfazl
    Yazdi, Mehran
    Owrangi, Amir M.
    Karbasi, Sareh
    Mosleh-Shirazi, Mohammad Amin
    BRACHYTHERAPY, 2022, 21 (06) : 933 - 942
  • [25] Feasibility of using software-aided selection of virtual monoenergetic level for optimal image quality of acute necrotising pancreatitis based on dual-energy computed tomography: a preliminary study
    Yuan Yuan
    Kai Liao
    Zixing Huang
    Liping Deng
    Hehan Tang
    Yi Wang
    Zheng Ye
    Xinyue Chen
    Bin Song
    Zhenlin Li
    BMC Medical Imaging, 23
  • [26] Feasibility of using software-aided selection of virtual monoenergetic level for optimal image quality of acute necrotising pancreatitis based on dual-energy computed tomography: a preliminary study
    Yuan, Yuan
    Liao, Kai
    Huang, Zixing
    Deng, Liping
    Tang, Hehan
    Wang, Yi
    Ye, Zheng
    Chen, Xinyue
    Song, Bin
    Li, Zhenlin
    BMC MEDICAL IMAGING, 2023, 23 (01)
  • [27] Dual-Energy Computed Tomography-How Accurate Is Gemstone Spectrum Imaging Metal Artefact Reduction? Its Application to Orthopedic Metal Implants
    Dabirrahmani, Dane
    Magnussen, John
    Appleyard, Richard Charles
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (06) : 925 - 935
  • [28] Accuracy of Pulmonary Nodule Volumetry Using Noise-Optimized Virtual Monoenergetic Image and Nonlinear Blending Image Algorithms in Dual-Energy Computed Tomography: A Phantom Study
    He, Changjiu
    Liu, Jieke
    Hu, Shibei
    Qing, Haomiao
    Qiao, Liang
    Luo, Hongbing
    Chen, Xiaoli
    Zhou, Peng
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2020, 44 (06) : 847 - 851
  • [29] Improving the Visualization of the Adrenal Veins Using Virtual Monoenergetic Images from Dual-Energy Computed Tomography before Adrenal Venous Sampling
    Wang, Yu
    Chen, Xiaohong
    Lu, Guoxiong
    Su, Yun
    Yang, Lingjie
    Shi, Guangzi
    Zhang, Fang
    Zhuo, Jiayi
    Duan, Xiaohui
    Hu, Huijun
    TOMOGRAPHY, 2023, 9 (02) : 485 - 496
  • [30] Determination of Renal Stone Composition in Phantom and Patients Using Single-Source Dual-Energy Computed Tomography
    Kulkarni, Naveen M.
    Eisner, Brian H.
    Pinho, Daniella F.
    Joshi, Mukta C.
    Kambadakone, Avinash R.
    Sahani, Dushyant V.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (01) : 37 - 45