AMPEROMETRIC ENZYME ELECTRODE FOR DETERMINATION OF THEOPHYLLINE IN SERUM

被引:16
|
作者
MCNEIL, CJ
COOPER, JM
SPOORS, JA
机构
[1] Department of Clinical Biochemistry, The Medical School, University of Newcastle upon Tyne, Newcastle upon Tyne
来源
BIOSENSORS & BIOELECTRONICS | 1992年 / 7卷 / 05期
关键词
ENZYME ELECTRODE; THEOPHYLLINE; SERUM; THEOPHYLLINE OXIDASE;
D O I
10.1016/0956-5663(92)85034-8
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper describes an amperometric enzyme electrode for the rapid determination of theophylline in serum. The method is based on the catalysed oxidation of theophylline by the haem-containing enzyme theophylline oxidase. Results are presented for two approaches. First, ferrocene monocarboxylic acid was used as a mediator. The second-order rate constant was 1.1 X 10(3) 1 mol-1 s-1. Secondly, the organic conducting salt NMP.TCNQ was used to construct enzyme electrodes. These electrodes were employed for the rapid (60 s) measurement of theophylline in serum at a working potential of +100 mV versus Ag/AgCl. Linear calibration curves were obtained over the clinically relevant range (y = 0.13x + 0.22, n = 8). Caffeine, theobromine and 3-methyl-xanthine at levels up to 100 mg l-1 do not interfere and 1-methylxanthine shows cross-reactivity at concentrations greater than 50 mg l-1.