EFFECTS OF MAGNETIC-RESONANCE-IMAGING ON IMPLANTABLE PERMANENT-MAGNETS

被引:0
|
作者
SCHNEIDER, ML
WALER, GB
DORMER, KJ
机构
[1] UNIV OKLAHOMA,SCH ELECT ENGN,OKLAHOMA CITY,OK 73112
[2] UNIV OKLAHOMA,HLTH SCI CTR,DEPT PHYSIOL,OKLAHOMA CITY,OK 73112
来源
AMERICAN JOURNAL OF OTOLOGY | 1995年 / 16卷 / 05期
关键词
D O I
暂无
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Implantable permanent magnets are increasingly used in devices for otolaryngologic applications. It is likely that at least some of the patients with implanted magnets will be in need of magnetic resonance imaging (MRI). The effect of an MRI scan on the magnetic properties of implanted permanent magnets has not been previously demonstrated. Some of the basic concepts and descriptive terminology used in industry regarding permanent magnets are reviewed. Experiments presented show that the MRI scan is capable of demagnetizing permanent magnets. A case history is also presented that demonstrates demagnetizing of an implanted Audiant magnet by an MRI scan.
引用
收藏
页码:687 / 689
页数:3
相关论文
共 50 条