DIFFERENTIAL-PULSE ADSORPTIVE STRIPPING VOLTAMMETRIC DETERMINATION OF SODIUM CROMOGLYCATE AT A HANGING MERCURY DROP ELECTRODE

被引:12
|
作者
MOREIRA, JC [1 ]
FOSTER, SE [1 ]
RODRIGUES, JA [1 ]
FOGG, AG [1 ]
机构
[1] LOUGHBOROUGH UNIV TECHNOL,DEPT CHEM,LOUGHBOROUGH LE11 3TU,LEICS,ENGLAND
关键词
SODIUM CROMOGLYCATE; ADSORPTIVE STRIPPING VOLTAMMETRY; HANGING MERCURY DROP ELECTRODE;
D O I
10.1039/an9921700989
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sodium cromoglycate is strongly adsorbed onto mercury, and can be determined readily at the 1 x 10(-10) mol dm-3 level, after adsorption onto a hanging mercury drop electrode from pH 4.5 acetate buffer solution. The adsorbed cromoglycate anion was reduced at -0.87 V versus Ag-AgCl, after accumulation at -0.6 V. Slightly higher signals can be obtained by accumulating at potentials more positive than the potential of zero charge (-0.5 V). Accumulation at -0.6 V, however, eliminates interference from more easily reduced adsorbates, and the analysis time is decreased. At the 0.5 x 10(-7) and 1 x 10(-7) mol dm-3 levels, peak height increased rectilinearly with accumulation time up to 5 and 3 min, respectively, and peaks of useful size were observed at accumulation times as short as 15 s. Relative standard deviations at the 1 X 10(-7) mol dm-3 level were typically <4% (five determinations). No significant interference was observed at this level from 1 mg dm-3 levels of several surfactants and 1 X 10(-6) mol dm-3 levels of several heavy metal ions. Human serum albumin at the 1.6 X 10(-7) mol dm-3 level caused the height of the peak to be reduced to 25% of its original size. The method was tested satisfactorily by analysing four formulation solutions of the drug.
引用
收藏
页码:989 / 991
页数:3
相关论文
共 50 条
  • [21] CATHODIC STRIPPING VOLTAMMETRIC DETERMINATION OF PENTAMIDINE ISETHIONATE AT A HANGING MERCURY DROP ELECTRODE
    ZANONI, MVB
    FOGG, AG
    [J]. ANALYST, 1993, 118 (09) : 1163 - 1166
  • [22] DIFFERENTIAL-PULSE POLAROGRAPHY AND ADSORPTIVE STRIPPING VOLTAMMETRIC DETERMINATION OF URIC-ACID
    GANDOUR, MA
    ENSAFABOULKASIM
    AMRALLAH, AH
    FARGHALY, OA
    [J]. JOURNAL OF THE INDIAN CHEMICAL SOCIETY, 1992, 69 (12) : 813 - 818
  • [23] Cathodic stripping voltammetric determination of ceftazidime in urine at a hanging mercury drop electrode
    Ferreira, VS
    Zanoni, MVB
    Fogg, AG
    [J]. MICROCHEMICAL JOURNAL, 1997, 57 (01) : 115 - 122
  • [24] Differential Pulse Adsorption Stripping Voltammetric Determination of Copper(II) with 2-Mercaptobenzimidazol at a Hanging Mercury-Drop Electrode
    Ali Asghar Ensafi
    Sharyar Abbasi
    Hassan Rahimi Mansour
    Iraje Mohammadpour Baltork
    [J]. Analytical Sciences, 2001, 17 : 609 - 612
  • [25] Differential pulse adsorption stripping voltammetric determination of copper(II) with 2-mercaptobenzimidazol at a hanging mercury-drop electrode
    Ensafi, AA
    Abbasi, S
    Mansour, HR
    Baltork, IM
    [J]. ANALYTICAL SCIENCES, 2001, 17 (05) : 609 - 612
  • [26] Indirect determination of lutetium by differential pulse anodic stripping voltammetry at a hanging mercury drop electrode
    Kumric, Ksenija R.
    Trtic-Petrovic, Tatjana M.
    Ignjatovic, Ljubisa M.
    Comor, Jozef J.
    [J]. CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2008, 6 (01): : 65 - 69
  • [27] ADSORPTIVE STRIPPING VOLTAMMETRY OF CHLORDIAZEPOXIDE AT THE HANGING MERCURY DROP ELECTRODE
    LORENZO, E
    HERNANDEZ, L
    [J]. ANALYTICA CHIMICA ACTA, 1987, 201 : 275 - 280
  • [28] ADSORPTIVE STRIPPING VOLTAMMETRY OF CLOTIAPINE AT A HANGING MERCURY DROP ELECTRODE
    LEJEUNE, R
    [J]. ANALYTICA CHIMICA ACTA, 1992, 256 (01) : 59 - 63
  • [29] ADSORPTIVE STRIPPING VOLTAMMETRIC BEHAVIOR OF COPPER(II) AT A HANGING MERCURY DROP ELECTRODE IN THE PRESENCE OF EXCESS OF IMIDAZOLE
    ERTAS, FN
    MOREIRA, JC
    FOGG, AG
    [J]. ANALYST, 1991, 116 (04) : 369 - 372
  • [30] Differential-pulse voltammetric determination of low μgl-1 cyanide levels using EDTA, Cu(II) and a hanging mercury drop electrode
    LaFuente, JMG
    Martínez, EF
    Pérez, JAV
    Fernández, SF
    Ordieres, AJM
    Uría, JES
    Sánchez, MLF
    Sanz-Medel, A
    [J]. ANALYTICA CHIMICA ACTA, 2000, 410 (1-2) : 135 - 142