An institutional experience of quality assurance of a treatment planning system on photon beam

被引:3
|
作者
Ozguven, Yildiray [1 ]
Yaray, Kadir [2 ]
Alkaya, Fadime [3 ]
Yucel, Birsen [1 ]
Soyuer, Serdar [2 ]
机构
[1] Cumhuriyet Univ, Fac Med, Dept Radiat Oncol, TR-58140 Sivas, Turkey
[2] Erciyes Univ, Fac Med, Dept Radiat Oncol, TR-38039 Kayseri, Turkey
[3] Medicana Hosp, Dept Radiat Oncol, TR-34520 Istanbul, Turkey
关键词
Treatment planning system; Quality assurance; Local dose deviation; Confidence limit; Photon beam;
D O I
10.1016/j.rpor.2013.10.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: The purpose of the present study is to show the application of the IAEA TRS-430 QA procedures of Eclipse (TM) v7.5 TPS for photon energies. In addition, the trends of the deviations found in the conducted tests were determined. Background: In the past, the lack of complete TPS QA procedures led to some serious accidents. So, QA in the radiotherapy treatment planning process is essential for determination of accuracy in the radiotherapy process and avoidance of treatment errors. Materials and methods: The calculations of TPS and measurements of irradiations of the treatment device were compared in the study. As a result, the local dose deviation values (delta(1) : central beam axis, delta(2): penumbra and build up region, delta(3) : inside field, delta(4) : outside beam edges, delta(50-90): beam fringe, RW50 : radiological width) and their confidence limit values (including systematic and random errors) were obtained. Results: The confidence limit values of delta(4) were detected to increase with expanding field size. The values of delta(1) and delta(3) of hard wedge were larger than open fields. The values of delta(2) and delta(50-90) of the inhomogeneity effect test were larger, especially than other tests of this study. The average deviation was showed to increase with the rise of the wedge angle. The values of delta(3) and delta(4) of lung irradiation were outside tolerance. Conclusions: The QA of TPS was done and it was found that there were no reservations in its use in patient treatment. The trend of the deviations is shown. (C) 2013 Greater Poland Cancer Centre. Published by Elsevier Urban & Partner Sp. z o.o. All rights reserved.
引用
收藏
页码:195 / 205
页数:11
相关论文
共 50 条
  • [1] An institutional experience: quality assurance of a treatment planning system on electron beam
    Yaray, Kadir
    Ozguven, Yildiray
    Alkaya, Fadime
    Yucel, Birsen
    Soyuer, Serdar
    [J]. TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2013, 28 (04): : 154 - 162
  • [2] Quality assurance for a treatment planning system in scanned ion beam therapy
    Jäkel, O
    Hartmann, GH
    Karger, CP
    Heeg, P
    Rassow, J
    [J]. MEDICAL PHYSICS, 2000, 27 (07) : 1588 - 1600
  • [3] Photon Beam Quality Assurance with the Integral Quality Monitor System: A Multicenter Study
    Veres, A.
    Farrokhish, M.
    DeMarco, J.
    de los Santos, L. Fong
    Heaton, R.
    Islam, M.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E315 - E315
  • [4] Is Routine Treatment Planning System Quality Assurance Necessary?
    Alaei, P.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 128 - +
  • [5] First Quality Assurance Implementation of Treatment Planning System for Pencil Beam Scanning Proton Radiotherapy
    Xiong, W.
    Yu, G.
    Kang, M.
    Huang, S.
    Lin, H.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [7] Quality Assurance and QMP for Brachytherapy Treatment Delivery and Treatment Planning System
    Chilukuri, M.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3776 - 3776
  • [8] IMRT QUALITY ASSURANCE USING A SECOND TREATMENT PLANNING SYSTEM
    Anjum, Muhammad Naeem
    Parker, William
    Ruo, Russell
    Aldahlawi, Ismail
    Afzal, Muhammad
    [J]. MEDICAL DOSIMETRY, 2010, 35 (04) : 274 - 279
  • [9] Quality assurance and quality control in operable breast carcinoma external beam radiotherapy treatment planning
    Stankovic, J
    Saric, M
    Borojevic, N
    RadosevicJelic, L
    [J]. BALKAN CONGRESS OF ONCOLOGY, 1996, : 177 - 183
  • [10] Commissioning and quality assurance-of the Pinnacle3 radiotherapy treatment planning system for external beam photons
    Bedford, JL
    Childs, PJ
    Hansen, VN
    Mosleh-Shirazi, MA
    Verhaegen, F
    Warrington, AP
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (903): : 163 - 176