LONG-TERM PERFORMANCE OF ENDOCARDIAL PACING LEADS

被引:56
|
作者
HELGUERA, ME [1 ]
MALONEY, JD [1 ]
PINSKI, SL [1 ]
WOSCOBOINIK, JR [1 ]
WILKOFF, BL [1 ]
CASTLE, LW [1 ]
机构
[1] CLEVELAND CLIN FDN,DEPT CARDIOL,CLEVELAND,OH 44195
来源
关键词
CARDIAC PACING; ARTIFICIAL PACING; ENDOCARDIAL PACING ELECTRODES; POLYURETHANE INSULATION; BIPOLAR ELECTRODES;
D O I
10.1111/j.1540-8159.1994.tb01351.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To assess the performance of endocardial pacemaker leads and to identify factors associated with structural lead failure, medical records of 2,611 endocardial pacing leads fin 1,518 patients) implanted between 1980 and 1991, having at least 1 month of follow-up, were reviewed. Leads without structural failure had normal function at the last follow-up date, or were discontinued for reasons other than structural failure (patient death, infection, dislodgment, lead-pacemaker incompatibility, operative complication, or abandonment by telemetry not related to failure). Leads with suspected structural failures were invasively or noninvasively disconnected because of clinical malfunction (loss of capture or sensing, oversensing, elevated thresholds, or skeletal muscular stimulation). Leads with verified structural failures met the criteria for suspected lead failure and also had a visible defect seen in the operating room or on chest roentgenograms, a change in the impedance interpreted by the physician as lead disruption, or a manufacturer's return product report that confirmed structural failure. Variables analyzed included patients' age and gender, paced chamber, venous access, insulation materials, fixation mechanism, coaxial design, polarity and different lead models. The cumulative lead survival at 5 and 10 years were 97.4% and 92.9%, respectively, for suspected failures; and 98.7% and 97.3%, respectively, for verified failures. Leads in older patients greater than or equal to 65 years old), and leads in atrial position had fewer verified failures IP = 0.014 and P = 0.007, respectively). Unipolar leads else fended to perform better according to the verified definition (P = 0.07). The lead Medtronic 4012 had more suspected IP < 0.05) and more verified failures (P < 0.01), the lead CPI 4010 had more verified failures (P < 0.05) than the entire group of ventricular leads. Conclusions: Endocardial pacing leads implanted in atrial position, and implanted in older patients (greater than or equal to 65 years old) seems to have better long-term survival. Some lead models (Medtronic 4012 and CPI 4010) had poor survival rates, that could not be explained by the analyzed variables. The expect ed performance of endocardial pacing leads varies according to how failure is defined.
引用
收藏
页码:56 / 64
页数:9
相关论文
共 50 条
  • [41] ACTIVE FIXATION OF ENDOCARDIAL PACING LEADS - THE PREFERRED METHOD OF PEDIATRIC PACING
    FRIEDMAN, RA
    MOAK, JP
    GARSON, A
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08): : 1213 - 1216
  • [42] ACUTE AND LONG-TERM MECHANICAL AND ELECTRO-PHYSIOLOGICAL CHARACTERISTICS OF TRANSVENOUS ENDOCARDIAL ATRIAL LEADS
    KRUSE, I
    RYDEN, L
    YDSE, B
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (03): : A55 - A55
  • [43] A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads
    Lie Liu
    Jiaojiao Tang
    Hu Peng
    Shulin Wu
    Chunying Lin
    Dongli Chen
    Qianhuan Zhang
    Yuanhong Liang
    Silin Chen
    Yan Chen
    Huiqiang Wei
    Scientific Reports, 5
  • [44] A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads
    Liu, Lie
    Tang, Jiaojiao
    Peng, Hu
    Wu, Shulin
    Lin, Chunying
    Chen, Dongli
    Zhang, Qianhuan
    Liang, Yuanhong
    Chen, Silin
    Chen, Yan
    Wei, Huiqiang
    SCIENTIFIC REPORTS, 2015, 5
  • [45] Long-term performance of transvenous, steroid-eluting, high impedance, passive-fixation ventricular pacing leads
    Yeh, KH
    Wang, CC
    Wen, MS
    Chou, CC
    Yeh, SJ
    Wu, DL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (10): : 1399 - 1404
  • [46] Long-term thresholds of nonsteroidal permanent pacing leads: A 5-year study
    Gumbrielle, TP
    Bourke, JP
    Sinkovic, M
    Tynan, M
    Kittpawong, P
    Gold, RG
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (05): : 829 - 835
  • [47] THROMBOEMBOLIC EVENTS IN LEFT VENTRICULAR ENDOCARDIAL PACING: LONG-TERM OUTCOMES FROM A MULTICENTRE UK REGISTRY
    Sawhney, Vinit
    Domenichini, Guilia
    Gamble, James
    Furniss, Guy
    Panagopoulos, Dimitrios
    Haywood, Guy
    Rajappan, Kim
    Campbell, Niall
    Lambiase, Pier
    Sporton, Simon
    Earley, Mark
    Lowe, Martin
    Dhinoja, Mehul
    Hunter, Ross
    Betts, Tim
    Schilling, Richard
    HEART, 2017, 103 : A30 - A30
  • [48] ACTIVELY ANCHORABLE ENDOCARDIAL LEADS FOR CARDIAC PACING IN INFANCY
    FRAEDRICH, G
    MULCH, J
    NETZ, H
    SCHELD, HH
    THORACIC AND CARDIOVASCULAR SURGEON, 1981, 29 : 16 - 16
  • [49] COMPARISON OF ACTIVE AND PASSIVE ADHERING LEADS FOR ENDOCARDIAL PACING
    FURMAN, S
    PANNIZZO, F
    CAMPO, I
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (04): : 417 - 427
  • [50] MODIFIED TECHNIQUE FOR LONG-TERM ENDOCARDIAL PACEMAKING
    GOTSMAN, MS
    BECK, W
    PILLER, LW
    BOSMAN, SCW
    BARNARD, CN
    SCHRIRE, V
    BRITISH MEDICAL JOURNAL, 1966, 2 (5526): : 1357 - &