DETERMINATION OF TRACE AMOUNTS OF COBALT BY FLOW-INJECTION WITH SPECTROPHOTOMETRIC DETECTION

被引:14
|
作者
PYRZYNSKA, K [1 ]
JANISZEWSKA, Z [1 ]
SZPUNARLOBINSKA, J [1 ]
TROJANOWICZ, M [1 ]
机构
[1] UNIV WARSAW, DEPT CHEM, PL-02093 WARSAW, POLAND
关键词
FLOW INJECTION; COBALT DETERMINATION; ONLINE PRECONCENTRATION; MODIFIED ALUMINA; SPECTROPHOTOMETRIC DETECTION;
D O I
10.1039/an9941901553
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection system incorporating a microcolumn with modified alumina was used in conjunction with spectrophotometric detection at 520 nm for the preconcentration and determination of cobalt. The analyte was preconcentrated on-line as the cobalt-nitroso-R-salt complex during the sampling stage and elution was effected by injection of sodium hydroxide solution. The limit of detection for a 10 min preconcentration time was 3.5 ng ml(-1) and the relative standard deviation at the 20 ng ml(-1) level was 3.1%, No significant interferences due to the presence of several metal ions was found. The procedure was applied to the determination of cobalt in oriental tobacco leaves and pyrite ore certified reference materials.
引用
收藏
页码:1553 / 1556
页数:4
相关论文
共 50 条