ULTRA MICRO GLUTAMATE SENSOR USING PLATINIZED CARBON-FIBER ELECTRODE AND INTEGRATED COUNTER ELECTRODE

被引:17
|
作者
TAMIYA, E [1 ]
SUGIURA, Y [1 ]
TAKEUCHI, T [1 ]
SUZUKI, M [1 ]
KARUBE, I [1 ]
AKIYAMA, A [1 ]
机构
[1] TOKYO INST TECHNOL,GRAD SCH NAGATSUTA,MIDORI KU,YOKOHAMA,KANAGAWA 227,JAPAN
关键词
D O I
10.1016/0925-4005(93)87003-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An integrated ultra micro glutamate sensor has been constructed with a 7 mum diameter platinized carbon-fiber disk (PCD) electrode and a platinum thin-film (PTF) counter electrode fabricated on the glass capillary tube. By platinization, the electrode activity of the carbon fiber is improved. In order to obtain a stable response, a pulse potential is applied for hydrogen peroxide measurement. The PTF counter electrode shows good stability and can be used as a substitute for a silver-silver chloride electrode. Since the integrated PCD electrode shows good characteristics as a hydrogen peroxide sensor, glutamate oxidase is immobilized onto the tip of the PCD electrode to construct the ultra micro glutamate sensor. The sensor shows stable response to glutamate and a response time within 12 s. The calibration range for glutamate measurement is 50-800 muM.
引用
收藏
页码:179 / 184
页数:6