A SIMPLE NUMERICAL METHOD TO COMPUTE THE SIGNAL-TO-NOISE RATIO OF A MAGNETIC RESONANCE IMAGING SURFACE COIL

被引:0
|
作者
Rojas, R. [1 ]
Rodriguez, A. O. [1 ]
机构
[1] Univ Autonoma Metropolitana Iztapalapa, Ctr Nacl Invest Instrumentac & Imagenol Med, Ave San Rafael Atlixco 186, Mexico City 09340, DF, Mexico
关键词
Electric fields - Electromagnetic field effects - Finite element method - Numerical methods - Electromagnetic fields - Magnetic resonance imaging - Maxwell equations - Phantoms;
D O I
10.2528/PIERM09040604
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The quality of a magnetic resonance image can be reliably measured by the signal-to-noise ratio. This widely accepted parameter is a function of the magnetic field generated by the coil and the electric field produced by the sample to be imaged. A simple numerical method is proposed to calculate the coil signal-to-noise ratio of a circular shaped coil and a spherical phantom. The phantom is composed of two concentric sphere simulating a brain-skull model. The electromagnetic fields produced were then numerically computed by solving Maxwell's equations with the finite element method implemented in a commercial software tool. The electric and magnetic fields were used to numerically determine the signal-to-noise ratio using the quasi-static approach. The numerical results demonstrated that this simple method is able to calculate the signal-to-noise ratio of surface coils with simple coil geometries involving a simulated phantom.
引用
收藏
页码:109 / 122
页数:14
相关论文
共 50 条
  • [31] Signal-to-Noise Ratio Enhancement Method for Ultra-Low Field Magnetic Resonance Imaging Based on Low-Frequency Surface Metamaterials
    Kong, Xiaohan
    Zhang, Yana
    Wu, Jiamin
    He, Yucheng
    Xu, Zheng
    [J]. Diangong Jishu Xuebao/Transactions of China Electrotechnical Society, 2024, 39 (13): : 3917 - 3927
  • [32] Maximizing signal-to-noise ratio in the presence of coil coupling
    Duensing, GR
    Brooker, HR
    Fitzsimmons, JR
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 111 (03): : 230 - 235
  • [33] CALCULATION OF THE SIGNAL-TO-NOISE RATIO FOR SIMPLE SURFACE COILS AND ARRAYS OF COILS
    WANG, JM
    REYKOWSKI, A
    DICKAS, J
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (09) : 908 - 917
  • [34] AN UNBIASED SIGNAL-TO-NOISE RATIO MEASURE FOR MAGNETIC-RESONANCE IMAGES
    MCGIBNEY, G
    SMITH, MR
    [J]. MEDICAL PHYSICS, 1993, 20 (04) : 1077 - 1078
  • [35] Single Image Signal-to-Noise Ratio Estimation for Magnetic Resonance Images
    K. S. Sim
    M. A. Lai
    C. P. Tso
    C. C. Teo
    [J]. Journal of Medical Systems, 2011, 35 : 39 - 48
  • [36] Single Image Signal-to-Noise Ratio Estimation for Magnetic Resonance Images
    Sim, K. S.
    Lai, M. A.
    Tso, C. P.
    Teo, C. C.
    [J]. JOURNAL OF MEDICAL SYSTEMS, 2011, 35 (01) : 39 - 48
  • [37] METHOD FOR DETERMINATION OF SIGNAL-TO-NOISE RATIO
    BURUNSUZ.ES
    LEONOV, OV
    [J]. RADIO ENGINEERING AND ELECTRONIC PHYSICS-USSR, 1970, 15 (11): : 1964 - &
  • [38] A COMPARISON OF SIGNAL-TO-NOISE LEVELS FROM SEVERAL RECEIVER COIL DESIGNS IN NUCLEAR-MAGNETIC-RESONANCE IMAGING
    LERSKI, RA
    MCROBBIE, DW
    STRAUGHAN, K
    JENNINGS, S
    ORR, JS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (704): : 820 - 820
  • [39] Signal-to-noise ratio, contrast-to-noise ratio and pharmacokinetic modeling considerations in dynamic contrast-enhanced magnetic resonance imaging
    Li, Xin
    Huang, Wei
    Rooney, William D.
    [J]. MAGNETIC RESONANCE IMAGING, 2012, 30 (09) : 1313 - 1322
  • [40] SIGNAL-TO-NOISE RATIO CONSIDERATIONS IN RADIOGRAPHIC IMAGING
    CHAKRABORTY, DP
    BARNES, GT
    [J]. MEDICAL PHYSICS, 1982, 9 (04) : 637 - 637