Dosimetric Study of Automatic Brain Metastases Planning in Comparison with Conventional Multi-Isocenter Dynamic Conformal Arc Therapy and Gamma Knife Radiosurgery for Multiple Brain Metastases

被引:21
|
作者
Mori, Yoshimasa [1 ]
Kaneda, Naoki [2 ]
Hagiwara, Masahiro [3 ]
Ishiguchi, Tuneo [1 ]
机构
[1] Aichi Med Univ, Dept Radiol & Radiat Oncol, Nagakute, Aichi, Japan
[2] Archie Med Univ, Dept Radiol & Radiat Oncol, Nagakute, Aichi, Japan
[3] Nagoya Kyoritsu Hosp, Nagoya Radiosurg Ctr, Nagoya, Aichi, Japan
来源
CUREUS | 2016年 / 8卷 / 11期
关键词
brain; metastasis; stereotactic radiosurgery; multiple lesions; gamma knife; linear accelerator; dynamic conformal arc; automatic brain metastases planning element; iplan; dose gradient;
D O I
10.7759/cureus.882
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective The efficacy of stereotactic radiosurgery (SRS) using Gamma Knife (GK) (Elekta, Tokyo) is well known. Recently, Automatic Brain Metastases Planning (ABMP) Element (BrainLAB, Tokyo) for a LINAC-based radiation system was commercially released. It covers multiple off-isocenter targets simultaneously inside a multi-leaf collimator field and enables SRS / stereotactic radiotherapy (SRT) with a single group of LINAC-based dynamic conformal multi-arcs (DCA) for multiple brain metastases. In this study, dose planning of ABMP (ABMP-single isocenter DCA (ABMP-SIDCA)) for SRS of small multiple brain metastases was evaluated in comparison with those of conventional multi-isocenter DCA (MIDCA-SRS) (iPlan, BrainLAB, Tokyo) and GK-SRS (GKRS). Methods Simulation planning was performed with ABMP-SIDCA and GKRS in the two cases of multiple small brain metastases (nine tumors in both), which had been originally treated with iPlan-MIDCA. First, a dosimetric comparison was done between ABMP-SIDCA and iPlan-MIDCA in the same setting of planning target volume (PTV) margin and D95 (dose covering 95% of PTV volume). Second, dosimetry of GKRS with a margin dose of 20 Gy was compared with that of ABMP-SIDCA in the setting of PTV margin of 0, 1 mm, and 2 mm, and D95=100% dose (20 Gy). Results First, the maximum dose of PTV and minimum dose of gross tumor volume (GTV) were significantly greater in ABMP-SIDCA than in iPlan-MIDCA. Conformity index (CI, 1/Paddick's CI) and gradient index (GI, V (half of prescription dose) / V (prescription dose)) in ABMP-SIDCA were comparable with those of iPlan-MIDCA. Second, PIV (prescription isodose volume) of GKRS was consistent with that of 1 mm margin - ABMP-SIDCA plan in Case 1 and that of no-margin ABMP-SIDCA plan in Case 2. Considering the dose gradient, the mean of V (half of prescription dose) of ABMP-SIDCA was not broad, comparable to GKRS, in either Case 1 or 2. Conclusions The conformity and dose gradient with ABMP-SIDCA were as good as those of conventional MIDCA for each lesion. If the conditions of the LINAC system permit a minimal PTV margin (1 mm or less), ABMP-SIDCA might provide excellent dose fall-off comparable with that of GKRS thereby enabling a short treatment time.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Dosimetric Comparison between Single-Isocenter RapidArc and Multi-Isocenter Circular Cone Arc for Radiosurgery to Multiple Brain Metastases
    Dai, P.
    Zhang, P.
    Luo, L.
    Deng, G.
    Huang, Q.
    Cai, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E508 - E508
  • [2] Dosimetric Evaluation of Single Versus Multi-isocenter Frameless Radiosurgery for Multiple Brain Metastases
    Camarata, A. S.
    Siddiqi, A. S.
    Prabhu, R. S.
    Ogunleye, T.
    Koontz-Raisig, W. G.
    Tejani, M.
    Curran, W. J.
    Crocker, I. R.
    Shu, H. G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S758 - S758
  • [3] Dosimetric study between a single isocenter dynamic conformal arc therapy technique and Gamma Knife radiosurgery for multiple brain metastases treatment: impact of target volume geometrical characteristics
    Michel Chea
    Karen Fezzani
    Julian Jacob
    Marguerite Cuttat
    Mathilde Croisé
    Jean-Marc Simon
    Loïc Feuvret
    Charles-Ambroise Valery
    Philippe Maingon
    Mohamed-Amine Benadjaoud
    Catherine Jenny
    [J]. Radiation Oncology, 16
  • [4] Dosimetric study between a single isocenter dynamic conformal arc therapy technique and Gamma Knife radiosurgery for multiple brain metastases treatment: impact of target volume geometrical characteristics
    Chea, Michel
    Fezzani, Karen
    Jacob, Julian
    Cuttat, Marguerite
    Croise, Mathilde
    Simon, Jean-Marc
    Feuvret, Loic
    Valery, Charles-Ambroise
    Maingon, Philippe
    Benadjaoud, Mohamed-Amine
    Jenny, Catherine
    [J]. RADIATION ONCOLOGY, 2021, 16 (01)
  • [5] Dosimetric comparison of inverse and forward planning for Gamma Knife stereotactic radiosurgery of brain metastases
    Xu, Qianyi
    Luo, Dershan
    Mulvihill, David
    Goldman, Warren
    Turtz, Alan
    Kubicek, Gregory J.
    [J]. MEDICAL DOSIMETRY, 2020, 45 (04) : 317 - 320
  • [6] Single Isocenter Radiosurgery for the Simultaneous Treatment of Multiple Brain Metastases: Volumetric Modulated Arc Therapy Or Dynamic Conformal Arc?
    Maurer, J.
    Sintay, B.
    Manning, M.
    Wiant, D.
    Liu, H.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3580 - 3580
  • [7] Plan comparison of volumetric modulated arc therapy with gamma knife radiosurgery for multiple brain metastases
    Nomoto, A. K.
    Sakumi, A.
    Takahashi, W.
    Igaki, H.
    Iwai, Y.
    Yoda, K.
    Nakagawa, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S220 - S221
  • [8] Dosimetric Comparison of Volumetric-Modulated Arc Therapy and Dynamic Conformal Arc Therapy in Three-fraction Single-isocenter Stereotactic Radiosurgery for Multiple Brain Metastases
    Turkkan, Gorkem
    Bilici, Nazli
    Sertel, Huseyin
    Tavli, Busra
    Ozkirim, Muge
    Fayda, Merdan
    [J]. TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2022, 37 (01): : 67 - 73
  • [9] Volumetric Arc Therapy (RapidArc) vs Gamma Knife Radiosurgery for Multiple Brain Metastases: Not Only a Dosimetric Issue
    Fiorentino, Alba
    Levra, Niccolo Giaj
    Mazzola, Rosario
    Fersino, Sergio
    Ricchetti, Francesco
    Alongi, Filippo
    [J]. NEUROSURGERY, 2015, 77 (02) : E310 - E310
  • [10] Stereotactic radiosurgery with single-isocenter dynamic arc therapy of multiple brain metastases
    Bodensohn, R.
    Kaempfel, A.
    Fleischmann, D. F.
    Hofmaier, J.
    Hadi, I.
    Garny, S.
    Reiner, M.
    Corradini, S.
    Forbrig, R.
    Thon, N.
    Belka, C.
    Niyazi, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S470 - S471