SQUARE-WAVE ANODIC-STRIPPING VOLTAMMETRY OF LEAD AND CADMIUM AT CYLINDRICAL GRAPHITE FIBER MICROELECTRODES WITH INSITU PLATED MERCURY FILMS

被引:30
|
作者
WOJCIECHOWSKI, M
BALCERZAK, J
机构
[1] Department of Chemistry and Biochemistry, University of Maryland, Baltimore, MD 21228, Baltimore County Campus
关键词
ANODIC STRIPPING VOLTAMMETRY; CADMIUM; LEAD;
D O I
10.1016/S0003-2670(00)83017-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Square-wave anodic stripping voltammetry (SWASV) at a cylindrical, 8-mu-m graphite fiber microelectrode with an in situ plated thin mercury film was investigated in oxygen-free and air-saturated solutions containing high and low concentrations of a supporting electrolyte. Owing to the non-planar (cylindrical) diffusion-enhanced mass transport at the microelectrode surface, very high steady-state current densities were observed at the microelectrode, e.g. 1.2 mA l mmol-1 cm-2 for a -1.0 V reduction of 0.59 mM Hg(II) in 0.1 M acetate buffer. To obtain such high current densities at a conventional rotating disk electrode (RDE), the RDE must be rotated at 1600 rpm. Because of the high speed and background rejection features of square-wave voltammetry, SWASV measurements (without background subtraction) of lead and cadmium at the in situ mercury-plated microelectrodes are not affected by the presence of dissolved oxygen, despite a high yield of oxygen diffusion to the microelectrode. Enhanced SWASV peaks were recorded for lead and cadmium in millimolar solutions of supporting electrolytes. In a non-deaerated and quiescent solution of 0.84 mM acetate buffer, for example, the method yielded a detection limit of 5 x 10(-10) M for each metal. This represents a very impressive sensitivity of the method, especially when one considers the very small dimensions of the microelectrode, the absence of solution stirring and purging and the relatively short deposition time used (2 min). In addition to improved sensitivity, SWASV at in situ mercury-plated microelectrodes offers other advantages over the commonly used anodic stripping methods such as differential-pulse anodic stripping voltammetry, particularly in terms of speed, convenience and applicability. Because of the elimination of both the oxygen removal step and solution stirring, and because of the use of the very fast stripping technique, the method is significantly faster and technically simpler, and thus more convenient for analysis. Also, because of the incorporation of cylindrical microelectrodes, the method does not require high concentrations of supporting electrolytes or sophisticated instrumentation, and can be applied for the analysis of microvolume samples.
引用
收藏
页码:433 / 445
页数:13
相关论文
共 50 条
  • [41] DETERMINATION OF LEAD AND CADMIUM IN URINE BY DIFFERENTIAL-PULSE ANODIC-STRIPPING VOLTAMMETRY
    ONAR, AN
    TEMIZER, A
    ANALYST, 1987, 112 (03) : 227 - 229
  • [42] Mercury-free analysis of lead in drinking water by anodic stripping square wave voltammetry
    Wilburn, Jeremy P.
    Brown, Kyle L.
    Cliffel, David E.
    JOURNAL OF CHEMICAL EDUCATION, 2007, 84 (02) : 312 - 314
  • [43] Square-wave stripping voltammetry with thin mercury-film ultramicroelectrodes
    Jin, JY
    Kinugasa, K
    Takeuchi, T
    Miwa, T
    BUNSEKI KAGAKU, 1998, 47 (06) : 349 - 354
  • [44] Simultaneous determination of cadmium, lead and copper in chocolate samples by square wave anodic stripping voltammetry
    Silva, Larissa Pinto
    Campos, Naira Da Silva
    Lisboa, Thalles Pedrosa
    de Faria, Lucas Vinicius
    Matos, Maria Auxiliadora Costa
    Matos, Renato Camargo
    de Sousa, Rafael Arromba
    FOOD ADDITIVES AND CONTAMINANTS PART A-CHEMISTRY ANALYSIS CONTROL EXPOSURE & RISK ASSESSMENT, 2021, 38 (03): : 418 - 426
  • [45] ANODIC-STRIPPING VOLTAMMETRY WITH FLORENCE MERCURY FILM ELECTRODE - DETERMINATION OF COPPER, LEAD AND CADMIUM IN SEA-WATER
    LUND, W
    SALBERG, M
    ANALYTICA CHIMICA ACTA, 1975, 76 (01) : 131 - 141
  • [46] Trace analysis of lead at a Nafion-modified electrode using square-wave anodic stripping voltammetry
    Crowley, K
    Cassidy, J
    ELECTROANALYSIS, 2002, 14 (15-16) : 1077 - 1082
  • [47] DIRECT DETERMINATION OF MERCURY IN SEA-WATER BY ANODIC-STRIPPING VOLTAMMETRY WITH A GRAPHITE ELECTRODE
    FUKAI, R
    HUYNHNGOC, L
    ANALYTICA CHIMICA ACTA, 1976, 83 (MAY) : 375 - 379
  • [48] THEORETICAL AND EXPERIMENTAL STUDY OF ANODIC STRIPPING VOLTAMMETRY WITH MERCURY-PLATED GRAPHITE ELECTRODES
    PERONE, SP
    BRUMFIEL.A
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1967, 13 (1-2): : 124 - &
  • [49] Inexpensive and disposable copper mini-sensor modified with bismuth for lead and cadmium determination using square-wave anodic stripping voltammetry
    Figueiredo-Filho, Luiz Carlos S.
    Janegitz, Bruno C.
    Fatibello-Filho, Orlando
    Marcolino-Junior, Luiz Humberto
    Banks, Craig E.
    ANALYTICAL METHODS, 2013, 5 (01) : 202 - 207
  • [50] Theory of Anodic Stripping Square Wave Voltammetry on Spherical Mercury Electrodes
    Komorsky-Lovric, Sebojka
    Lovric, Milivoj
    CROATICA CHEMICA ACTA, 2014, 87 (03) : 287 - 290