ENDLESS-LOOP TACHYCARDIAS - DESCRIPTION AND 1ST CLINICAL-RESULTS OF A NEW FULLY-AUTOMATIC PROTECTION ALGORITHM

被引:14
|
作者
NITZSCHE, R
GUEUNOUN, M
LAMAISON, D
LASCAULT, G
PIOGER, G
RICHARD, M
MALHERBE, O
LIMOUSIN, M
机构
[1] CTR HOSP REG & UNIV,HOSP NORD,MARSEILLE,FRANCE
[2] CTR HOSP REG & UNIV,CLERMONT FERRAND,FRANCE
[3] HOP JEAN ROSTAND,IVRY,FRANCE
[4] CLIN ALLERAY,PARIS,FRANCE
[5] CTR HOSP ST MAIO,ST MALO,FRANCE
来源
关键词
DDD PACING; ENDLESS-LOOP TACHYCARDIA; RETROGRADE CONDUCTION;
D O I
10.1111/j.1540-8159.1990.tb06877.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Endless-loop tachycardia (ELT) is one of the most common pacemaker mediated tachycardia. An innovative ELT protection algorithm has proven to be clinically effective. A new improved version that will eliminate the need to program any parameter is now under clinical evaluation. Nine patients entered the study: six men and three women, aged 52 +/- 22 years. This automatic algorithm needs only 10 cycles to detect and confirm an ELT. Three hundred thirty-three ELTs lasting more than 9 cycles have been induced and analyzed. The total results are the following: mean duration: 6.7 sec +/- 3.1; mean ELT rate: 137 +/- 21.9 bpm, mean programmed upper rate limit (URL): 142.5 +/- 26.5 bpm (Only 70% of ELTs presented rates equal to programmed URL). (1) ELTs reduced by postventricular atrial refractory period (PVARP) extension on one cycle: 291 ELTs (87%). ELT rate: 128.5 +/- 18.2 bpm. (2) Retrograde block: algorithm operation may induce a retrograde block due to a short atrioventricular delay (AVD) applied during the confirmation phase to discriminate an ELT from a stable sinus rhythm. Thirty-two ELTs (10%) have been reduced and detected on a retrograde block occurrence. (3) Algorithm failure due to an unstable ventriculoatrial conduction time (VACT) even at fixed rate or to a retrograde Wenckebach behavior on AVD reduction during the confirmation phase. A total of 10 algorithms failed to detect or confirm an ELT have been recorded (3%). Mean duration: 8.2 +/- 4.2 sec, mean ELT rate: 148.9 +/- 14.3 bpm. This new fully automatic algorithm has reduced 97% of ELTs, including high rate episodes (100-175 bpm). It allows 1:1 tracking adapted to the needs of the patient, by programming a short PVARP and a physiological AVD.
引用
收藏
页码:1711 / 1718
页数:8
相关论文
共 50 条
  • [41] 1ST CLINICAL-RESULTS OF INTRAOPERATIVE APPLICATION OF STREPTOKINASE-STREPTODORNASE IN CHILDREN
    MEIER, H
    DIETL, KH
    WILLITAL, GH
    [J]. LANGENBECKS ARCHIV FUR CHIRURGIE, 1985, 366 : 191 - 193
  • [42] A NEW INTERDISCIPLINARY TREATMENT REGIMEN FOR ADVANCED HEAD AND NECK TUMORS - RATIONALE, FEASIBILITY AND 1ST CLINICAL-RESULTS
    SCHMITT, G
    HIGI, M
    STUPP, H
    SCHERER, E
    [J]. STRAHLENTHERAPIE, 1983, 159 (08) : 470 - 473
  • [43] MICROWAVE HYPERTHERMIA CONTROLLED BY MICROWAVE RADIOMETRY - TECHNICAL ASPECTS AND 1ST CLINICAL-RESULTS
    CHIVE, M
    PLANCOT, M
    GIAUX, G
    PREVOST, B
    [J]. JOURNAL OF MICROWAVE POWER AND ELECTROMAGNETIC ENERGY, 1984, 19 (04) : 233 - 241
  • [44] SPIRAL CT - DIFFERENTIAL-DIAGNOSIS OF CERVICAL TUMEFACTIONS - 1ST CLINICAL-RESULTS
    STEINKAMP, HJ
    KNOBBER, D
    HOSTEN, N
    FELIX, R
    [J]. LARYNGO-RHINO-OTOLOGIE, 1993, 72 (07) : 361 - 367
  • [45] BIOMECHANICAL PROPERTIES AND 1ST CLINICAL-RESULTS OF A NEWLY DESIGNED INTRAMEDULLARY FIXATION DEVICE
    HAAS, N
    KRETTEK, C
    GOTZEN, L
    [J]. JOURNAL OF BIOMECHANICS, 1985, 18 (07) : 544 - 544
  • [46] INTRAVENOUS PROUROKINASE THERAPY IN ACUTE MYOCARDIAL-INFARCTION - 1ST CLINICAL-RESULTS
    KASPER, W
    MEINERTZ, T
    HOHNLOSER, S
    ENGLER, H
    HASLER, C
    JUST, H
    [J]. ZEITSCHRIFT FUR KARDIOLOGIE, 1986, 75 : 49 - 49
  • [47] RADIOGRAPHY OF THE CARPAL SCAPHOID - EXPERIMENTAL EVALUATION OF THE CARPAL BOX AND 1ST CLINICAL-RESULTS
    TIELVANBUUL, MMC
    VANBEEK, EJR
    DIJKSTRA, PF
    BAKKER, ADJ
    GRIFFIOEN, FMM
    BROEKHUIZEN, TH
    [J]. INVESTIGATIVE RADIOLOGY, 1992, 27 (11) : 954 - 959
  • [48] THE NEW ASTROLABE OF SANTIAGO (CHILE) - DESCRIPTION OF THE INSTRUMENT AND 1ST RESULTS
    CHOLLET, F
    NOEL, F
    [J]. ASTRONOMY & ASTROPHYSICS, 1993, 276 (02) : 655 - 662
  • [49] A NEW LASER SCANNING SYSTEM FOR VIDEO OPHTHALMOSCOPY - 1ST CLINICAL-RESULTS WITH AND WITHOUT DIGITAL IMAGE-PROCESSING
    FABIAN, E
    MERTZ, M
    HOFMANN, H
    WERTHEIMER, R
    FOOS, C
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1990, 197 (01) : 55 - 63
  • [50] A NEW LASER SCANNING SYSTEM FOR VIDEO OPHTHALMOSCOPY - 1ST CLINICAL-RESULTS WITH AND WITHOUT DIGITAL IMAGE-PROCESSING
    FABIAN, E
    MERTZ, M
    HOFMANN, H
    WERTHEIMER, R
    FOOS, C
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1990, 196 (06) : 489 - 494