USE OF A LARGE AREA PHOTODIODE IN CSI(TL) SCINTILLATION-COUNTERS

被引:20
|
作者
GUNJI, S
HANADA, N
HOSHINO, T
KAMAE, T
MIYAZAKI, S
TAKAHASHI, T
TAMURA, T
TANAKA, M
YAMAGAMI, T
MURAKAMI, H
MORI, K
TANAKA, H
YAMAMOTO, K
机构
[1] INST SPACE & ASTRONAUT SCI,SAGAMIHARA 229,JAPAN
[2] RIKKYO UNIV,DEPT PHYS,TOSHIMA KU,TOKYO 171,JAPAN
[3] CLEAR PULSE KK,OHTA KU,TOKYO 143,JAPAN
[4] HAMAMATSU PHOTON KK,DIV SOLID STATE,HAMAMATSU 435,JAPAN
关键词
D O I
10.1016/0168-9002(90)90719-M
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A newly developed large area photodiode has been studied as the photon sensor for medium-size CsI(Tl) scintillation counters. The photodiode has a sensitive area of 18 × 18 mm2 and a thickness of 500 μm. When combined with a CsI(Tl) crystal of 2.5 × 2.5 × 5.0 cm3, it gives an energy resolution of around 5.1% (FWHM) for 1.33 MeV gamma-rays. Characteristics of the low noise preamplifier designed for the CsI(Tl)-photodiode counter will also be given. © 1990.
引用
收藏
页码:400 / 404
页数:5
相关论文
共 50 条
  • [21] A Study of CsI(Tl) Scintillator with Optimized Conditions of Large Area Avalanche Photodiode
    Kim, Minjeong
    Kang, Heedong
    Kim, H. J.
    Kim, Wan
    Park, H.
    Kim, Sunghwan
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 586 - 589
  • [22] ANSI STANDARDS FOR LIQUID SCINTILLATION-COUNTERS
    FERRIS, R
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1980, 31 (01): : 10 - 10
  • [23] INVESTIGATION OF AMPLITUDE CHARACTERISTICS OF SCINTILLATION-COUNTERS
    GLAVANAKOV, IV
    STIBUNOV, VN
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1978, 21 (01) : 32 - 35
  • [24] COMPARISON OF REFLECTORS USED IN SCINTILLATION-COUNTERS
    SALOMON, M
    NUCLEAR INSTRUMENTS & METHODS, 1978, 153 (01): : 289 - 290
  • [25] STABILIZATION OF PHOTOMULTIPLIERS IN LIQUID SCINTILLATION-COUNTERS
    SOINI, E
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (08): : 980 - 984
  • [26] CsI(Tl) plus photodiode scintillation assemblies for γ-ray and proton detectors
    Kudin, A. M.
    Borodenko, Yu. A.
    Grinyov, B. V.
    Didenko, A. V.
    Dudnik, A. V.
    Zaslavsky, B. G.
    Valtonen, E.
    Eronen, T.
    Peltonen, J.
    Lehti, J.
    Kettunen, H.
    Virtanen, A.
    Huovelin, J.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2010, 53 (01) : 39 - 44
  • [27] SCINTILLATION-COUNTERS OF 100CM LENGTH
    ASTVATSATUROV, RG
    IVANOV, VI
    KNAPIK, E
    MALAKHOV, AI
    KHACHATURYAN, MN
    PRIBORY I TEKHNIKA EKSPERIMENTA, 1975, (06): : 49 - 51
  • [28] LARGE-AREA SCINTILLATION COUNTERS
    BATCHELOR, R
    SMITH, JW
    BLACKMORE, L
    NATURE, 1950, 165 (4200) : 688 - 688
  • [29] PERFORMANCE OF LARGE AREA SCINTILLATION COUNTERS
    BARNABY, CF
    BARTON, JC
    PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON, 1960, 76 (491): : 745 - 753
  • [30] INFLUENCE OF QUENCHING ON THE BACKGROUND OF LIQUID SCINTILLATION-COUNTERS
    ALESSIO, M
    ALLEGRI, L
    BELLA, F
    IMPROTA, S
    NUCLEAR INSTRUMENTS & METHODS, 1978, 157 (03): : 579 - 582