SPLIT SHIELD FOR CORRECTION OF RADIATION ERRORS IN STEADY-STATE THERMAL-CONDUCTIVITY MEASUREMENTS

被引:0
|
作者
TALPE, J
AVENDANO, A
机构
[1] Laboratorio de Temperaturas Intermedias, Facultad de Ciencias Exactas, Universidad Nacional de Salta
关键词
THERMAL CONDUCTIVITY; SPLIT THERMAL SHIELDS; RADIATION ERRORS;
D O I
10.1016/0011-2275(93)90198-W
中图分类号
O414.1 [热力学];
学科分类号
摘要
An axial heat flow, steady state method for thermal conductivity measurements above 80 K on poor conductors is described. A split shield (one part held at the sink temperature and the other kept at the hot-end temperature) is shown to be successful for the correction of radiation errors. Measurements on poly(methyl methacrylate) (PMMA) are reported.
引用
收藏
页码:833 / 835
页数:3
相关论文
共 50 条
  • [1] STEADY-STATE THERMAL-CONDUCTIVITY MEASUREMENTS OF A1N AND SIC SUBSTRATE MATERIALS
    DETTMER, ES
    ROMENESKO, BM
    CHARLES, HK
    CARKHUFF, BG
    MERRILL, DJ
    [J]. IEEE TRANSACTIONS ON COMPONENTS HYBRIDS AND MANUFACTURING TECHNOLOGY, 1989, 12 (04): : 543 - 547
  • [2] THE STEADY-STATE THERMAL-CONDUCTIVITY OF BRICK WALLS - A METHOD OF MEASUREMENT
    GAVIOLI, G
    TENAGLIA, A
    [J]. TRANSACTIONS AND JOURNAL OF THE BRITISH CERAMIC SOCIETY, 1982, 81 (02): : 52 - 56
  • [3] Thermal-conductivity apparatus for steady-state, comparative measurement of ceramic coatings
    Slifka, AJ
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2000, 105 (04) : 591 - 605
  • [4] STEADY-STATE EFFECTIVE THERMAL-CONDUCTIVITY OF CLOSED WETTED POROUS SYSTEMS
    PALOSAARI, SM
    CORNISH, ARH
    [J]. ACTA POLYTECHNICA SCANDINAVICA-CHEMICAL TECHNOLOGY SERIES, 1975, (125): : 3 - 34
  • [5] SOME OVERLOOKED ERRORS IN THERMAL-CONDUCTIVITY MEASUREMENTS
    BECK, AE
    [J]. GEOPHYSICS, 1975, 40 (01) : 173 - 174
  • [6] EXPERIMENTAL-DETERMINATION OF THE EFFECTIVE THERMAL-CONDUCTIVITY OF GLASS BY STEADY-STATE METHOD
    ENDRYS, J
    BLAZEK, A
    EDEROVA, J
    [J]. GLASTECHNISCHE BERICHTE-GLASS SCIENCE AND TECHNOLOGY, 1993, 66 (6-7): : 151 - 157
  • [7] REDUCTION OF RADIATION ERRORS IN THERMAL-CONDUCTIVITY MEASUREMENTS ABOVE 100-K
    TALPE, JH
    BEKERIS, VI
    SANCHEZ, MJ
    SIMONELLI, G
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (04): : 1054 - 1056
  • [8] Steady-state thermal conductivity measurements of super-hard materials
    Khorunzhii, I
    Gabor, HB
    Job, R
    Fahrner, WR
    Denisenko, A
    Brunner, D
    Peschek, U
    [J]. MEASUREMENT, 2002, 32 (03) : 163 - 172
  • [9] Thermal conductivity of magnesium oxide from absolute, steady-state measurements
    Slifka, AJ
    Filla, BJ
    Phelps, JM
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1998, 103 (04) : 357 - 363
  • [10] STEADY-STATE HOT-WIRE THERMAL-CONDUCTIVITY SET-UP FOR FLUIDS
    VILCU, R
    CIOCHINA, A