Thermal-conductivity apparatus for steady-state, comparative measurement of ceramic coatings

被引:14
|
作者
Slifka, AJ [1 ]
机构
[1] Natl Inst Stand & Technol, Mat Sci & Engn Lab, Mat Reliabil Div, Boulder, CO 80303 USA
关键词
guarded hot plate; infrared microscope; thermal; barrier coating; thermal conductivity; yttria-stabilized-zirconia;
D O I
10.6028/jres.105.050
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An apparatus has been developed to measure the thermal conductivity of ceramic coatings. Since the method uses an infrared microscope for temperature measurement, coatings as thin as 20 mu m can, in principle, be measured using this technique. This steady-state, comparative measurement method uses the known thermal conductivity of the substrate material as the reference material for heat-flow measurement. The experimental method is validated by measuring a plasma-sprayed coating that has been previously measured using an absolute, steady-state measurement method. The new measurement method has a relative standard uncertainty of about 10 %. The measurement of the plasma-sprayed coating gives 0.58 W.m(-1).K-1 which compares well with the 0.62 W.m(-1).K-1 measured using the absolute method.
引用
收藏
页码:591 / 605
页数:15
相关论文
共 50 条
  • [1] A steady-state Bi-substrate technique for measurement of the thermal conductivity of ceramic coatings
    Tan, J. C.
    Tsipas, S. A.
    Golosnoy, I. O.
    Curran, J. A.
    Paul, S.
    Clyne, T. W.
    [J]. SURFACE & COATINGS TECHNOLOGY, 2006, 201 (3-4): : 1414 - 1420
  • [2] THE STEADY-STATE THERMAL-CONDUCTIVITY OF BRICK WALLS - A METHOD OF MEASUREMENT
    GAVIOLI, G
    TENAGLIA, A
    [J]. TRANSACTIONS AND JOURNAL OF THE BRITISH CERAMIC SOCIETY, 1982, 81 (02): : 52 - 56
  • [3] A Steady-State Apparatus to Measure the Thermal Conductivity of Solids
    Giovanni A. Longo
    [J]. International Journal of Thermophysics, 2008, 29 : 664 - 677
  • [4] MEASUREMENT OF THE THERMAL-CONDUCTIVITY OF UNFROZEN AND FROZEN FOOD MATERIALS BY A STEADY-STATE METHOD WITH COAXIAL DUAL-CYLINDER APPARATUS
    PONGSAWATMANIT, R
    MIYAWAKI, O
    YANO, T
    [J]. BIOSCIENCE BIOTECHNOLOGY AND BIOCHEMISTRY, 1993, 57 (07) : 1072 - 1076
  • [5] A steady-state apparatus to measure the thermal conductivity of solids
    Longo, Giovanni A.
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (02) : 664 - 677
  • [6] An improved steady-state apparatus for measuring thermal conductivity of soils
    Alrtimi, A.
    Rouainia, M.
    Manning, D. A. C.
    [J]. INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2014, 72 : 630 - 636
  • [7] APPARATUS FOR DETERMINING THE THERMAL-CONDUCTIVITY OF CERAMIC MATERIALS
    VED, VE
    IVANOV, VA
    LUSHPENKO, SF
    [J]. INDUSTRIAL LABORATORY, 1992, 58 (11): : 1062 - 1066
  • [8] Design and construction of a new steady-state apparatus for medium thermal conductivity measurement at high temperature
    Wang, Yong
    Xiao, Peng
    Dai, Jingmin
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (10):
  • [9] STEADY-STATE EFFECTIVE THERMAL-CONDUCTIVITY OF CLOSED WETTED POROUS SYSTEMS
    PALOSAARI, SM
    CORNISH, ARH
    [J]. ACTA POLYTECHNICA SCANDINAVICA-CHEMICAL TECHNOLOGY SERIES, 1975, (125): : 3 - 34
  • [10] DESIGN OF A STEADY-STATE, PARALLEL-PLATE THERMAL CONDUCTIVITY APPARATUS FOR NANOFLUIDS AND COMPARATIVE MEASUREMENTS WITH TRANSIENT HWTC APPARATUS
    Kostic, Milivoje M.
    Walleck, Casey J.
    [J]. PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION - 2010, VOL 5, PTS A AND B, 2012, : 1457 - 1464