BREAST THICKNESS IN ROUTINE MAMMOGRAMS - EFFECT ON IMAGE QUALITY AND RADIATION-DOSE

被引:66
|
作者
HELVIE, MA
CHAN, HP
ADLER, DD
BOYD, PG
机构
[1] Department of Radiology, Taubman Center, University of Michigan Medical Ctr., Ann Arbor, MI 48109-0326
关键词
D O I
10.2214/ajr.163.6.7992731
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare the thickness of the compressed breast between mediolateral oblique and craniocaudal mammograms and to relate these differences in thickness to image quality and radiation dose. These differences may partially explain why some subtle tumors are better visualized on the cranio caudal view. SUBJECTS AND METHODS. The study population consisted of 250 paired mediolateral oblique and craniocaudal mammograms obtained on one mammographic unit by seven certified mammography technologists during a P-month period. Only women with breast implants,prior lumpectomy and radiotherapy, or chest wall deformity were excluded. The digital readout of compressed breast thickness and applied compression force was recorded. Mammographic positioning was assessed using standard criteria. Absorbed radiation dose at different thicknesses was measured with a BR-12 breast phantom. Image quality differences for geometric unsharpness and contrast were calculated for the observed breast thickness differences between mediolateral oblique and craniocaudal mammograms. RESULTS. The mean thickness of the compressed breast on the craniocaudal view was less than the mean thickness on the mediolateral oblique view (4.4 versus 4.8 cm, p < .0001) despite the greater force used to compress the breast for mediolateral oblique than for craniocaudal views (93 versus 86 newtons, p < .0001). The breast thickness on the mediolateral oblique view exceeded that on the craniocaudal view in 98 (84%) of 117 pairs that differed in thickness by 5 mm or more and 46 (94%) of 49 pairs that differed by 10 mm or more (p < .0001). Geometric unsharpness increased by 8% and 19% when a 4.4-cm-thick breast was compared to a 4.8- and 5.4-cm-thick breast, respectively. A 5% and 12% loss of contrast was noted when a 4.4-cm-thick breast was compared to a 4.8- and 5.4-cm-thick breast, Mean glandular radiation dose at 4.4, 4.8, and 5.4 cm was 1.40, 1.70, and 2.33 mGy, respectively. CONCLUSION, The compressed breast is 8% thicker on mediolateral oblique than on craniocaudal mammograms, a small but statistically significant difference. This difference results in a small loss of spatial and contrast resolution on the mediolateral oblique views and an increase in radiation dose. These image quality differences may partially explain why some subtle carcinomas are better visualized on the craniocaudal view.
引用
收藏
页码:1371 / 1374
页数:4
相关论文
共 50 条
  • [21] Image quality in mammograms with low compressed breast thicknesses
    Loveland, J.
    Mackenzie, A.
    17TH INTERNATIONAL WORKSHOP ON BREAST IMAGING, IWBI 2024, 2024, 13174
  • [22] Effect of additional filtration on radiation dose and image quality
    Gai, Xinghui
    Yang, Dewu
    Li, Xunru
    Sui, Yan
    Liu, Kang
    Wang, Junying
    Zhao, Mengjuan
    2022 IEEE 6TH ADVANCED INFORMATION TECHNOLOGY, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IAEAC), 2022, : 1807 - 1811
  • [23] RADIATION-DOSE AND 2ND BREAST-CANCER
    BASCO, VE
    COLDMAN, AJ
    ELWOOD, JM
    YOUNG, MEJ
    BRITISH JOURNAL OF CANCER, 1985, 52 (03) : 319 - 325
  • [24] SIMPLIFIED ABSORBED FRACTION METHOD FOR RADIATION-DOSE CALCULATIONS IN ROUTINE CLINICAL SITUATIONS
    CHANDRA, R
    RADIOLOGY, 1974, 111 (02) : 415 - 416
  • [25] EFFECT OF THE RADIATION-DOSE ON THE EFFICIENCY OF VOLUME POLYETHYLENE CHARGE
    PROKOPEV, OV
    ROZNO, AG
    GROMOV, VV
    ZHURNAL FIZICHESKOI KHIMII, 1988, 62 (03): : 821 - 823
  • [26] Strategies for Radiation-dose Reduction and Image-quality Optimization in Multidetector Computed Tomographic Coronary Angiography
    Torres, Felipe S.
    Crean, Andrew M.
    Nguyen, Elsie T.
    Paul, Narinder
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2010, 61 (05): : 271 - 279
  • [27] RELATIONSHIP BETWEEN RADIATION DOSE AND IMAGE QUALITY IN DIGITAL BREAST TOMOSYNTHESIS
    Alakhras, Maram M.
    Mello-Thoms, Claudia
    Bourne, Roger
    Rickard, Mary
    Diffey, Jennifer
    Brennan, Patrick C.
    RADIATION PROTECTION DOSIMETRY, 2017, 173 (04) : 351 - 360
  • [28] RADIATION-DOSE DURING HYSTEROSALPINGOGRAPHY USING A DIGITAL IMAGE INTENSIFICATION TECHNIQUE
    HAUSSLER, MD
    MERK, J
    TOMCZAK, R
    SCHNARKOWSKI, P
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1992, 157 (06): : 596 - 598
  • [29] Multidetector CT of bronchiectasis: Effect of radiation dose on image quality
    Yi, CA
    Lee, KS
    Kim, TS
    Han, DH
    Sung, YM
    Kim, S
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (02) : 501 - 505
  • [30] EFFECTIVE RADIATION-DOSE FOR LOCAL CONTROL OF ADVANCED BREAST-CANCER
    KEYS, HM
    RUBIN, P
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1976, : 92 - 92