MODEL OF THE MECHANISM OF LIQUID-METAL ION-SOURCE EMISSION

被引:1
|
作者
PTITSIN, VE
机构
[1] Institute for Analytical Instrumentation, Russian Academy of Sciences, 198103, St. Petersburg
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1994年 / 65卷 / 04期
关键词
D O I
10.1063/1.1144957
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The results of calculations of the liquid metal subsurface layer atom binding energy and temperature under operating conditions of a Ga liquid metal ion source (LMIS) are given. The concepts developed in this article can help to explain the presence of excited neutral atoms of the LMIS substance in the desorption flux.