RESPIRATORY MECHANICS BY LEAST-SQUARES FITTING IN MECHANICALLY VENTILATED PATIENTS - APPLICATIONS DURING PARALYSIS AND DURING PRESSURE SUPPORT VENTILATION

被引:95
|
作者
IOTTI, GA [1 ]
BRASCHI, A [1 ]
BRUNNER, JX [1 ]
SMITS, T [1 ]
OLIVEI, M [1 ]
PALO, A [1 ]
VERONESI, R [1 ]
机构
[1] HAMILTON BONADUZ AG, BONADUZ, SWITZERLAND
关键词
RESPIRATORY MECHANICS; RESPIRATORY RESISTANCE; RESPIRATORY COMPLIANCE; MECHANICAL VENTILATION; PRESSURE SUPPORT VENTILATION;
D O I
10.1007/BF01707409
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate a least squares fitting technique for the purpose of measuring total respiratory compliance (C-rs) and resistance (R(rs)) in patients submitted to partial ventilatory support, without the need for esophageal pressure measurement. Design: Prospective, randomized study. Setting: A general ICU of a University Hospital. Patients. 11 patients in acute respiratory failure, intubated and assisted by pressure support ventilation (PSV). Interventions: Patients were ventilated at 4 different levels of pressure support. At the end of the study, they were paralyzed for diagnostic reasons and submitted to volume controlled ventilation (CMV). Measurements and results: A least squares fitting (LSF) method was applied to measure C-rs and R(rs) at different levels of pressure support as well as in CMV. C-rs and R(rs) calculated by the LSF method were compared to reference values which were obtained in PSV by measurement of esophageal pressure, and in CMV by the application of the constant flow end-inspiratory occlusion method. Inspiratory activity was measured by P-0.1. In CMV, C-rs and R(rs) measured by the LSF method are close to quasistatic compliance (-1.5 +/- 1.5 ml/cmH(2)O) and to the mean value of minimum and maximum end-inspiratory resistance (+0.9 +/- 2.5 cmH(2)O/(l/s)). Applied during PSV, the LSF method leads to gross underestimation of R(rs) (-10.4 +/- 2.3 cmH(2)O/(l/s)) and overestimation of C-rs (+35.2 +/- 33 ml/cmH(2)O) whenever the set pressure support level is low and the activity of the respiratory muscles is high (P-0.1 was 4.6 +/- 3.1 cmH(2)O). However, satisfactory estimations of C-rs and R(rs) by the LSF method were obtained at increased pressure support levels, resulting in a mean error of - 0.4 +/- 6 ml/cmH(2)O and -2.8 +/- 1.5 cmH(2)O/(l/s), respectively. This condition was coincident with a P-0.1 of 1.6 +/- 0.7 cmH(2)O. Conclusion: The LSF method allows non-invasive evaluation of respiratory mechanics during PSV, provided that a near-relaxation condition is obtained by means of an adequately increased pressure support level. The measurement of P-0.1 may be helpful for titrating the pressure support in order to obtain the condition of near-relaxation.
引用
收藏
页码:406 / 413
页数:8
相关论文
共 50 条
  • [41] Flow triggering added to pressure support ventilation improves comfort and reduces work of breathing in mechanically ventilated patients
    Barrera, R
    Melendez, J
    Ahdoot, M
    Huang, Y
    Leung, D
    Groeger, JS
    JOURNAL OF CRITICAL CARE, 1999, 14 (04) : 172 - 176
  • [42] Patient-ventilator interaction and sleep in mechanically ventilated patients: Pressure support versus proportional assist ventilation
    Bosma, Karen
    Ferreyra, Gabriela
    Ambrogio, Cristina
    Pasero, Daniela
    Mirabella, Lucia
    Braghiroli, Alberto
    Appendini, Lorenzo
    Mascia, Luciana
    Ranieri, V. Marco
    CRITICAL CARE MEDICINE, 2007, 35 (04) : 1048 - 1054
  • [43] Pressure support ventilation mode versus pressure support ventilation+T-piece trial as a weaning modality in mechanically ventilated patients with chronic obstructive pulmonary disease
    Ahmed M. Azouz
    Rafat T. El-Sokary
    Hassan A. El-Latif
    Ghada Ahmed
    Egyptian Journal of Bronchology, 2019, 13 : 87 - 92
  • [44] Impact Of Synchronization With Spontaneous Breathing During Pressure Controlled Modes In Mechanically Ventilated Patients
    Rittayamai, N.
    Beloncle, F.
    Chen, L.
    Waheed, R.
    Richard, J-C. M.
    Brochard, L.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2015, 191
  • [45] Changes in pulse pressure variability during cardiac resynchronization therapy in mechanically ventilated patients
    Cornelius Keyl
    Jochem Stockinger
    Sven Laule
    Klaus Staier
    Jochen Schiebeling-Römer
    Christoph Wiesenack
    Critical Care, 11
  • [46] Changes in pulse pressure variability during cardiac resynchronization therapy in mechanically ventilated patients
    Keyl, Cornelius
    Stockinger, Jochem
    Laule, Sven
    Staier, Klaus
    Schiebeling-Roemer, Jochen
    Wiesenack, Christoph
    CRITICAL CARE, 2007, 11 (02)
  • [47] The respiratory system during resuscitation:: a review of the history, risk of infection during assisted ventilation, respiratory mechanics, and ventilation strategies for patients with an unprotected airway
    Wenzel, V
    Idris, AH
    Dörges, V
    Nolan, JP
    Parr, MJ
    Gabrielli, A
    Stallinger, A
    Lindner, KH
    Baskett, PJF
    RESUSCITATION, 2001, 49 (02) : 123 - 134
  • [48] Noninvasive method for measuring respiratory system compliance during pressure support ventilation
    Zhao, Zhanqi
    Eger, Marcus
    Handzsuj, Thomas
    Ranieri, V. Marco
    Appendini, Lorenzo
    Micelli, Cosimo
    Moeller, Knut
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 3808 - 3811
  • [49] A novel method for noninvasive estimation of respiratory effort during pressure support ventilation
    Pasillas-Lepine, William
    Tuffet, Samuel
    Soussen, Charles
    Gendreau, Segolene
    Boujelben, Mohamed Ahmed
    Mekonsto-Dessap, Armand
    Carteaux, Guillaum
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 93
  • [50] Acute effects of hyperoxemia on dyspnoea and respiratory variables during pressure support ventilation
    Volta, CA
    Alvisi, V
    Bertacchini, S
    Marangoni, E
    Ragazzi, R
    Verri, M
    Alvisi, R
    INTENSIVE CARE MEDICINE, 2006, 32 (02) : 223 - 229