RETENTION IN COUPLED CAPILLARY COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY WITH LIPIDS AS MODEL COMPOUNDS

被引:7
|
作者
KARLSSON, L
JAREMO, M
EMILSSON, M
MATHIASSON, L
JONSSON, JA
机构
[1] Department of Analytical Chemistry, University of Lund, Lund, 221 00
[2] Karlshamns Oils and Fats AB, Karlshamn
关键词
SUPERCRITICAL FLUID CHROMATOGRAPHY; COUPLED CAPILLARY COLUMNS; RETENTION PREDICTION; LIPIDS;
D O I
10.1007/BF02272256
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have investigated to which extent retention data, acquired on single capillary columns, can be used for predicting retention factors in a coupled column system. For this purpose we utilized a model mixture of 18 lipid components with widely different vapor pressures and polarities. The sample was chromatographed on two columns, SB-biphenyl-30 (70% methyl-30% biphenylpolysiloxane) and SB-cyanopropyl-50 (50% methyl-50% cyanopropylsiloxane). Experimental retention factors, acquired in coupled column systems with two columns connected in different order, were thus compared with values calculated from runs on each single column. The agreement between calculated and experimental values generally was better than 5% without any pressure drop correction. To study the possibility of predicting retention behavior in a wide pressure range from a limited number of experiments, we also investigated the relation between solute retention and mobile phase density. We found that all data could be fitted to second order equations, which gives the possibility to optimize the resolution with respect to pressure from a limited number of runs at different pressures.
引用
收藏
页码:402 / 410
页数:9
相关论文
共 50 条
  • [41] PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY COUPLED WITH ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY
    SADOUN, F
    VIRELIZIER, H
    ARPINO, PJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1993, 647 (02): : 351 - 359
  • [42] SIMPLE RESTRICTORS FOR CAPILLARY COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY
    GREEN, S
    BERTSCH, W
    [J]. JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1988, 11 (05): : 414 - 415
  • [43] Microelectrochemical detection in capillary column supercritical fluid chromatography
    Wallenborg, SR
    Markides, KE
    Nyholm, L
    [J]. PROCEEDINGS OF THE SYMPOSIUM ON CHEMICAL AND BIOLOGICAL SENSORS AND ANALYTICAL ELECTROCHEMICAL METHODS, 1997, 97 (19): : 665 - 673
  • [44] CHARACTERIZATION OF FUELS BY MULTIDIMENSIONAL SUPERCRITICAL-FLUID CHROMATOGRAPHY AND SUPERCRITICAL-FLUID CHROMATOGRAPHY MASS-SPECTROMETRY
    ANDERSSON, PE
    DEMIRBUKER, M
    BLOMBERG, LG
    [J]. JOURNAL OF CHROMATOGRAPHY, 1993, 641 (02): : 347 - 355
  • [45] ANALYSIS OF CARBAMATE PESTICIDES BY PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY
    BERGER, TA
    WILSON, WH
    DEYE, JF
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1994, 32 (05) : 179 - 184
  • [46] BIOANALYTICAL APPLICATION OF MULTIDIMENSIONAL OPEN TUBULAR COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY
    XIE, LQ
    MARKIDES, KE
    LEE, ML
    HOLLENBERG, NK
    WILLIAMS, GH
    GRAVES, SW
    [J]. CHROMATOGRAPHIA, 1993, 35 (7-8) : 363 - 371
  • [47] SOME THERMODYNAMIC ASPECTS OF PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY
    SHANG, DY
    GRANDMAISON, JL
    KALIAGUINE, S
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1994, 672 (1-2) : 185 - 201
  • [48] MECHANISM OF SOLUTE RETENTION IN CAPILLARY SUPERCRITICAL FLUID CHROMATOGRAPHY
    YONKER, CR
    FRYE, SL
    UDSETH, HR
    WRIGHT, BW
    SMITH, RD
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1985, 190 (SEP): : 73 - FUL
  • [49] AUTOMATED LARGE-VOLUME INJECTION IN CAPILLARY SUPERCRITICAL-FLUID CHROMATOGRAPHY
    BERG, BE
    GREIBROKK, T
    [J]. HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1994, 17 (01): : 49 - 50
  • [50] EFFECT OF PRESSURE ON RETENTION IN SUPERCRITICAL-FLUID CHROMATOGRAPHY WITH PACKED-COLUMNS
    SCHOENMAKERS, PJ
    VERHOEVEN, FCCJG
    [J]. JOURNAL OF CHROMATOGRAPHY, 1986, 352 : 315 - 328