MEASURING THE PREDICTIVE PERFORMANCE OF COMPUTER-CONTROLLED INFUSION PUMPS

被引:414
|
作者
VARVEL, JR
DONOHO, DL
SHAFER, SL
机构
[1] UNIV CALIF BERKELEY,DEPT STAT,BERKELEY,CA 94720
[2] STANFORD UNIV,MED CTR,DEPT ANESTHESIA,STANFORD,CA 94305
来源
关键词
ANESTHESIA; COMPUTER-ASSISTED EQUIPMENT; INFUSION PUMP; PERFORMANCE PHARMACOKINETICS; PREDICTIVE PERFORMANCE;
D O I
10.1007/BF01143186
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Current measures of the performance of computer-controlled infusion pumps (CCIPs) are poorly defined, of little use to the clinician using the CCIP, and pharmacostatistically incorrect. We propose four measures be used to quantitate the performance of CCIPs: median absolute performance error (MDAPE), median performance error (MDPE), divergence, and wobble. These measures offer several significant advantages over previous measures. First, their definitions are based on the performance error as a fraction of the predicted (rather than measured) drug concentration, making the measures much more useful to the clinician. Second, the measures are defined in a way that addresses the pharmacostatistical issue of appropriate estimation of population parameters. Finally, the measure of inaccuracy, MDAPE, is defined in a way that is consistent with iteratively reweighted least squares nonlinear regression, a commonly used method of estimating pharmacokinetic parameters. These measures make it possible to quantitate the overall performance of a CCIP or to compare the predictive performance of CCIPs which differ in either general approach (e.g., compartmental model driven vs. plasma efflux approach), pump mechanics, software algorithms, or pharmacokinetic parameter sets.
引用
收藏
页码:63 / 94
页数:32
相关论文
共 50 条
  • [1] TESTING COMPUTER-CONTROLLED INFUSION PUMPS BY SIMULATION
    SHAFER, SL
    SIEGEL, LC
    COOKE, JE
    SCOTT, JC
    ANESTHESIOLOGY, 1988, 68 (02) : 261 - 266
  • [2] Identification and targeting policies for computer-controlled infusion pumps
    Gentilini, A
    Frei, CW
    Glattfelder, AH
    Morari, M
    Schnider, TW
    CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 2000, 28 (1-2) : 179 - 185
  • [3] Food and drug administration regulations and computer-controlled infusion pumps
    Bazaral, MG
    Ciarkowski, A
    INTERNATIONAL ANESTHESIOLOGY CLINICS, 1995, 33 (03) : 45 - 63
  • [4] Predictive performance of computer-controlled infusion of remifentanil during propofol/remifentanil anaesthesia
    Mertens, MJ
    Engbers, FHM
    Burm, AGL
    Vuyk, J
    BRITISH JOURNAL OF ANAESTHESIA, 2003, 90 (02) : 132 - 141
  • [5] COMPUTER-CONTROLLED INFUSION
    SUTTMANN, H
    JUHL, G
    ANAESTHESIST, 1988, 37 (08): : 558 - 561
  • [6] PUMPSIM - A SOFTWARE PACKAGE FOR SIMULATING COMPUTER-CONTROLLED DRUG INFUSION PUMPS
    HAMANN, GE
    DOYLE, DJ
    JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (01): : 73 - 76
  • [7] THE HYBRID MODEL - A NEW PHARMACOKINETIC MODEL FOR COMPUTER-CONTROLLED INFUSION PUMPS
    WADA, DR
    WARD, DS
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (02) : 134 - 142
  • [8] Target-controlled infusion (TCI):: Administration of intravenous anaesthetics by means of computer-controlled infusion pumps
    Bruhn, J
    Röpcke, H
    Bouillon, T
    ANASTHESIOLOGIE & INTENSIVMEDIZIN, 2002, 43 (09): : 547 - 555
  • [9] COMPUTER-CONTROLLED INFUSION OF PROPOFOL
    TACKLEY, RM
    LEWIS, GTR
    PRYSROBERTS, C
    BOADEN, RW
    DIXON, J
    HARVEY, JT
    BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (01) : 46 - 53
  • [10] A general solution for nonparametric control of a linear system using computer-controlled infusion pumps
    Verotta, D
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (01) : 44 - 50