INDIRECT OPHTHALMOSCOPES - EVALUATION FOR POTENTIAL HAZARD

被引:6
|
作者
MILLER, SA [1 ]
JAMES, RH [1 ]
SLINEY, DH [1 ]
机构
[1] USA,ENVIRONM HYG AGCY,ABERDEEN PROVING GROUND,MD 21010
来源
APPLIED OPTICS | 1992年 / 31卷 / 10期
关键词
INDIRECT OPHTHALMOSCOPE; HAZARD ANALYSIS; OPTICAL RADIATION;
D O I
10.1364/AO.31.001631
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The optical radiation emitted from eight indirect ophthalmoscopes was measured. The levels were compared with the threshold limit values (TLVs) published by the American Conference of Governmental Industrial Hygienists. Many tested instruments were found to emit unnecessarily high levels of ultraviolet and infrared radiation in addition to visible light. The levels of optical radiation emitted from most of the ophthalmoscopes were below the applicable TLVs. Nevertheless, suitable blocking filters can reduce the emission of ultraviolet and infrared radiation. Use of the ophthalmoscopes at the lowest possible power setting and the shortest possible exposure time would further minimize patient exposure.
引用
收藏
页码:1631 / 1637
页数:7
相关论文
共 50 条