INDIRECT OPHTHALMOSCOPES - EVALUATION FOR POTENTIAL HAZARD

被引:6
|
作者
MILLER, SA [1 ]
JAMES, RH [1 ]
SLINEY, DH [1 ]
机构
[1] USA,ENVIRONM HYG AGCY,ABERDEEN PROVING GROUND,MD 21010
来源
APPLIED OPTICS | 1992年 / 31卷 / 10期
关键词
INDIRECT OPHTHALMOSCOPE; HAZARD ANALYSIS; OPTICAL RADIATION;
D O I
10.1364/AO.31.001631
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The optical radiation emitted from eight indirect ophthalmoscopes was measured. The levels were compared with the threshold limit values (TLVs) published by the American Conference of Governmental Industrial Hygienists. Many tested instruments were found to emit unnecessarily high levels of ultraviolet and infrared radiation in addition to visible light. The levels of optical radiation emitted from most of the ophthalmoscopes were below the applicable TLVs. Nevertheless, suitable blocking filters can reduce the emission of ultraviolet and infrared radiation. Use of the ophthalmoscopes at the lowest possible power setting and the shortest possible exposure time would further minimize patient exposure.
引用
收藏
页码:1631 / 1637
页数:7
相关论文
共 50 条
  • [1] ERECTING PRISM ATTACHMENT FOR INDIRECT OPHTHALMOSCOPES
    SCHIRMER, K
    SHEA, M
    ARCHIVES OF OPHTHALMOLOGY, 1970, 83 (05) : 529 - &
  • [2] SPECTRAL IRRADIANCES OF AND MAXIMAL PERMISSIBLE EXPOSURES TO 2 INDIRECT OPHTHALMOSCOPES
    KOSSOL, J
    COLE, C
    DAYHAWBARKER, P
    AMERICAN JOURNAL OF OPTOMETRY AND PHYSIOLOGICAL OPTICS, 1983, 60 (07): : 616 - 621
  • [3] EVALUATION OF POTENTIAL CARCINOGENIC HAZARD
    CARR, CJ
    KOLBYE, AC
    ADVANCES IN CHEMISTRY SERIES, 1985, (210): : 335 - 340
  • [4] EVALUATION OF POTENTIAL CARCINOGENIC HAZARD
    KOLBYE, AC
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1984, 187 (APR): : 81 - ENVR
  • [5] Development and preliminary results of a chromatic illumination system for indirect ophthalmoscopes
    Oliveira, Thiago Bellini
    Trevelin, Luis Carlos
    Araujo Moreira, Fernando Manoel
    Bagnato, Vanderlei Salvador
    Schor, Paulo
    Vieira de Carvalho, Luis Alberto
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2009, 72 (02) : 146 - 151
  • [6] Hazard analysis for the evaluation of the hazard potential of chemical process units
    Eichendorf, Klaus
    Guntrum, Eberhard
    Jochum, Christian
    Niemitz, Klaus-Jurgen
    2002, Wiley-VCH Verlag (25):
  • [7] Hazard analysis for the evaluation of the hazard potential of chemical process units
    Eichendorf, K
    Guntrum, E
    Jochum, C
    Niemitz, KJ
    CHEMICAL ENGINEERING & TECHNOLOGY, 2002, 25 (04) : 439 - 442
  • [8] EVALUATION OF POTENTIAL HAZARD OF PROPANIL TO GROUNDWATER
    TREVISAN, M
    MONTEPIANI, C
    GHEBBIONI, C
    DELRE, AAM
    CHEMOSPHERE, 1991, 22 (07) : 637 - 643
  • [9] COMPARISON OF PHOTOCOAGULATION WITH THE ARGON, KRYPTON, AND DIODE-LASER INDIRECT OPHTHALMOSCOPES IN RABBIT EYES
    BENNER, JD
    HUANG, M
    MORSE, LS
    HJELMELAND, LM
    LANDERS, MB
    OPHTHALMOLOGY, 1992, 99 (10) : 1554 - 1563
  • [10] Evaluation of the potential hazard of lanthanides to freshwater microcrustaceans
    Blinova, I
    Lukjanova, A.
    Muna, M.
    Vija, H.
    Kahru, A.
    SCIENCE OF THE TOTAL ENVIRONMENT, 2018, 642 : 1100 - 1107