COMPARISON OF RESISTANCE VERSUS THERMODYNAMIC TEMPERATURE OF PLATINUM RESISTANCE THERMOMETERS WITH THE ITS-90

被引:6
|
作者
CONNOLLY, JJ
JONES, TP
TAPPING, J
机构
[1] CSIRO Division of Applied Physics, National Measurement Laboratory, Lindfield
关键词
Block Body - International Temperature Scale - Metrology - Platinum Resistance Thermometers - Thermodynamic Temperatures;
D O I
10.1088/0026-1394/27/2/006
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Data relating the resistance of high temperature platinum resistance thermometers with thermodynamic temperatures, which were independent from those used to derive ITS-90, have been compared with ITS-90. Within the accuracy of the photoelectric pyrometry involved, ±0,40°C at the freezing point of silver, the measured thermodynamic temperatures agreed with the temperatures obtained using the ITS-90. In measurements performed during a 1,5 year period the typical random uncertainties, evidenced by variations in the resistance of platinum resistance thermometers at the triple point of water, were equivalent to ±0,015°C at the freezing point of silver, whilst, in this period, the drifts in the measured values of the resistance ratio at the freezing point of silver were equivalent to 0,03°C.