EFFECTS OF CUFF INFLATION ON SELF-RECORDED BLOOD-PRESSURE

被引:35
|
作者
VEERMAN, DP
VANMONTFRANS, GA
WIELING, W
机构
[1] Department of Internal Medicine, Academic Medical Centre, University of Amsterdam
来源
LANCET | 1990年 / 335卷 / 8687期
关键词
D O I
10.1016/0140-6736(90)90676-V
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Changes in continuously recorded 'Finapres' finger blood pressure in ten normotensive and seven hypertensive subjects induced by self-inflation of the cuff or just wearing the inflated cuff were studied. Inflating the cuff caused an instantaneous rise in systolic blood pressure of 13 and 12 mm Hg (hypertensive and normotensive subjects, respectively). Wearing the inflated cuff did not change blood pressure. Thus the rise in pressure was related to the muscular activity required for cuff inflation. Systolic blood pressure took on average 7 s and at most 21 s to return to baseline level after stopping cuff inflation. Since first Korotkoff sounds may already be heard after 10-15 s when following recommended procedures, self-recorded systolic blood pressure may be recorded as too high when subjects inflate their cuff at too low a pressure or deflate it too fast. © 1990.
引用
收藏
页码:451 / 453
页数:3
相关论文
共 50 条
  • [1] EVIDENCE OF RISE IN BLOOD-PRESSURE DUE TO INFLATING THE CUFF - EFFECTS ON SELF-RECORDED BLOOD-PRESSURE
    ECKERT, S
    GLEICHMANN, S
    OHLMEIER, H
    SCHMIDT, HK
    GLEICHMANN, U
    [J]. NIEREN-UND HOCHDRUCKKRANKHEITEN, 1993, 22 (07) : 297 - 300
  • [2] INFLATING ONES OWN CUFF DOES NOT INCREASE SELF-RECORDED BLOOD-PRESSURE
    VEERMAN, DP
    VANMONTFRANS, GA
    KAREMAKER, JM
    WIELING, W
    [J]. JOURNAL OF HYPERTENSION, 1988, 6 : S77 - S78
  • [3] AN EVALUATION OF SELF-RECORDED BLOOD-PRESSURE DURING DRUG TRIALS
    GOULD, BA
    HORNUNG, RS
    KIESO, H
    CASHMAN, PMM
    RAFTERY, EB
    [J]. HYPERTENSION, 1986, 8 (04) : 267 - 271
  • [4] HOME MEASUREMENT WITH CONFIDENCE - DISTRIBUTION OF SELF-RECORDED BLOOD-PRESSURE IN SUBJECTS WITH LOW BLOOD-PRESSURE
    KUYCK, WGE
    MONTFRANS, GAV
    [J]. JOURNAL OF HYPERTENSION, 1989, 7 : S378 - S378
  • [5] CROSSOVER DESIGN TO TEST ANTIHYPERTENSIVE DRUGS WITH SELF-RECORDED BLOOD-PRESSURE
    MENARD, J
    SERRURIER, D
    BAUTIER, P
    PLOUIN, PF
    CORVOL, P
    [J]. HYPERTENSION, 1988, 11 (02) : 153 - 159
  • [6] Reference values for self-recorded blood pressure.
    Thijs, L
    Staessen, JA
    Celis, H
    deGaudemaris, R
    Imai, Y
    Julius, S
    Fagard, R
    [J]. HYPERTENSION, 1997, 30 (04) : 81 - 81
  • [7] Self-recorded blood pressures
    Vidt, DG
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1998, 158 (05) : 425 - 425
  • [8] THE EFFECT OF BLOOD-PRESSURE CUFF INFLATION ON SLEEP - A POLYSOMNOGRAPHIC EXAMINATION
    DIMSDALE, JE
    COY, TV
    ANCOLIISRAEL, S
    CLAUSEN, J
    BERRY, CC
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 1993, 6 (10) : 888 - 891
  • [9] Reference values for self-recorded blood pressure - A meta-analysis of summary data
    Thijs, L
    Staessen, JA
    Celis, H
    de Gaudemaris, R
    Imai, Y
    Julius, S
    Fagard, R
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1998, 158 (05) : 481 - 488
  • [10] RELIABILITY OF SELF-RECORDED BEHAVIORS
    SIMKINS, L
    [J]. BEHAVIOR THERAPY, 1971, 2 (01) : 83 - 87