Virtual fluoroscopy for intraoperative C-arm positioning and radiation dose reduction

被引:14
|
作者
De Silva, Tharindu [1 ]
Punnoose, Joshua [1 ]
Uneri, Ali [1 ]
Mahesh, Mahadevappa [2 ]
Goerres, Joseph [1 ]
Jacobson, Matthew [1 ]
Ketcha, Michael D. [1 ]
Manbachi, Amir [1 ]
Vogt, Sebastian [3 ]
Kleinszig, Gerhard [3 ]
Khanna, Akhil Jay [4 ]
Wolinksy, Jean-Paul [5 ]
Siewerdsen, Jeffrey H. [1 ,2 ,5 ]
Osgood, Greg [4 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21218 USA
[3] Siemens Healthineers, Erlangen, Germany
[4] Johns Hopkins Univ, Orthopaed Surg, Baltimore, MD USA
[5] Johns Hopkins Univ, Dept Neurosurg, Baltimore, MD 21218 USA
关键词
3D-2D registration; image-guided surgery; orthopedic surgery; digitally reconstructed radiographs; virtual fluoroscopy;
D O I
10.1117/1.JMI.5.1.015005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Positioning of an intraoperative C-arm to achieve clear visualization of a particular anatomical feature often involves repeated fluoroscopic views, which cost time and radiation exposure to both the patient and surgical staff. A system for virtual fluoroscopy (called FluoroSim) that could dramatically reduce time-and dosespent "fluoro-hunting" by leveraging preoperative computed tomography (CT), encoded readout of C-arm gantry position, and automatic 3D-2D image registration has been developed. The method is consistent with existing surgical workflow and does not require additional tracking equipment. Real-time virtual fluoroscopy was achieved via mechanical encoding of the C-arm motion, C-arm geometric calibration, and patient registration using a single radiograph. The accuracy, time, and radiation dose associated with C-arm positioning were measured for FluoroSim in comparison with conventional methods. Five radiology technologists were tasked with acquiring six standard pelvic views pertinent to sacro-illiac, anterior-inferior iliac spine, and superior-ramus screw placement in an anthropomorphic pelvis phantom using conventional and FluoroSim approaches. The positioning accuracy, exposure time, number of exposures, and total time for each trial were recorded, and radiation dose was characterized in terms of entrance skin dose and in-room scatter. The geometric accuracy of FluoroSim was measured to be 1.6 +/- 1.1 mm. There was no significant difference (p > 0.05) observed in the accuracy or total elapsed time for C-arm positioning. However, the total fluoroscopy time required to achieve the desired view decreased by 4.1 s (4.7 +/- 3.6 s for conventional, compared with 0.5 +/- 0.0 s for FluoroSim, p < 0.05), and the total number of exposures reduced by 4.0 (6.4 +/- 4.8 for conventional, compared with 2.0 +/- 0.0 for FluoroSim, p < 0.05). These reductions amounted to a 50% to 78% decrease in patient entrance skin dose and a 55% to 70% reduction in in-room scatter. FluoroSim was found to reduce the radiation exposure required in C-arm positioning without diminishing positioning time or accuracy, providing a potentially valuable tool to assist technologists and surgeons. (c) 2018 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Measurements of Surgeons' Exposure to Ionizing Radiation Dose During Intraoperative Use of C-Arm Fluoroscopy
    Lee, Kisung
    Lee, Kyoung Min
    Park, Moon Seok
    Lee, Boram
    Kwon, Dae Gyu
    Chung, Chin Youb
    [J]. SPINE, 2012, 37 (14) : 1240 - 1244
  • [2] C-arm Positioning Using Virtual Fluoroscopy for Image-Guided Surgery
    De Silva, T.
    Punnoose, J.
    Uneri, A.
    Goerres, J.
    Jacobson, M.
    Ketcha, M. D.
    Manbachi, A.
    Vogt, S.
    Kleinszig, G.
    Khanna, A. J.
    Wolinksy, J. -P.
    Osgood, G.
    Siewerdsen, J. H.
    [J]. MEDICAL IMAGING 2017: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2017, 10135
  • [3] Utility of Intraoperative C-Arm Fluoroscopy for Frontal Sinusotomy
    Lee, Seung-Sin
    [J]. EWHA MEDICAL JOURNAL, 2009, 32 (02): : 61 - 64
  • [4] Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study
    Cha, Hyemi
    Lee, Kisung
    Park, Moon Seok
    Lee, Kyoung Min
    Cho, Kyeyoung
    Sung, Ki Hyuk
    [J]. JOURNAL OF RADIATION RESEARCH, 2020, 61 (05) : 705 - 711
  • [5] Radiation protection for an intraoperative X-ray source compared to C-arm fluoroscopy
    Schneider, Frank
    Clausen, Sven
    Jahnke, Anika
    Steil, Volker
    Bludau, Frederic
    Suetterlin, Marc
    Obertacke, Udo
    Wenz, Frederik
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (03): : 243 - 251
  • [6] C-ARM FLUOROSCOPY
    COHEN, MD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (03) : 646 - 647
  • [7] Intraoperative assessment of hindfoot alignment using C-arm fluoroscopy
    Park, Young Uk
    Lee, Jong Wha
    Chung, Jun Young
    Choi, Wan-Sun
    Kim, Taehun
    Seo, Young Wook
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2024, 29 (05) : 1248 - 1254
  • [8] Survey of terminology used for the intraoperative direction of C-arm fluoroscopy
    Pally, Elliott
    Kreder, Hans J.
    [J]. CANADIAN JOURNAL OF SURGERY, 2013, 56 (02) : 109 - 112
  • [9] Intraoperative C-Arm Radiation Affecting Factors and Reduction by an Intervention Program
    Bar-On, Elhanan
    Weigl, Daniel Martin
    Becker, Tali
    Katz, Kalman
    Konen, Osnat
    [J]. JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2010, 30 (04) : 320 - 323
  • [10] C-ARM FLUOROSCOPY - REPLY
    DAVIS, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (03) : 647 - 647