ACCURACY, REPRODUCIBILITY, AND VARIABILITY OF PORTABLE PEAK FLOWMETERS

被引:20
|
作者
JACKSON, AC [1 ]
机构
[1] BOSTON UNIV, DEPT BIOMED ENGN, RESP RES LAB, BOSTON, MA USA
关键词
ATS STANDARDS; NATIONAL ASTHMA EDUCATION PROGRAM RECOMMENDATIONS; PEAK FLOW MEASUREMENTS;
D O I
10.1378/chest.107.3.648
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The accuracy, reproducibility, and variability of seven commercially available peak flowmeters were assessed using the pulmonary waveform generator as recommended by the American Thoracic Society. The standard models were tested for peak flows of 100, 200, 320, 500, and 700 L/min. The low-range models were tested for peak flows of 80, 150, 200, 250, and 320 L/min. Most of the units provided highly repeatable estimates of peak now. However, the accuracy of several devices (Assess [Healthscan Products, Cedar Grove, NJ] and Mini-Wright [Clement Clarke International, Harlow, Essex, UK] low-range models, and the Ferraris [Ferraris Medical, Holland, NY] and Mini-Wright standard-range models) did not meet the National Asthma Education Program (NAEP) recommendations. The device with the best performance in terms of accuracy, variability, and reproducibility, and the only one in which all units tested met the NAEP recommendations, was the Astech (Center Laboratories, Port Washington, NY) full-range model.
引用
收藏
页码:648 / 651
页数:4
相关论文
共 50 条
  • [1] EVALUATION OF ACCURACY AND REPRODUCIBILITY OF PEAK FLOWMETERS AT 1,400 M
    GARDNER, RM
    CRAPO, RO
    JACKSON, BR
    JENSEN, RL
    [J]. CHEST, 1992, 101 (04) : 948 - 952
  • [2] THE USE OF PORTABLE PEAK FLOWMETERS IN THE SURVEILLANCE OF OCCUPATIONAL ASTHMA
    HENNEBERGER, PK
    STANBURY, MJ
    TRIMBATH, LS
    KIPEN, HM
    [J]. CHEST, 1991, 100 (06) : 1515 - 1521
  • [3] AN EVALUATION OF THE ACCURACY OF ASSESS AND MINIWRIGHT PEAK FLOWMETERS
    SHAPIRO, SM
    HENDLER, JM
    OGIRALA, RG
    ALDRICH, TK
    SHAPIRO, MB
    [J]. CHEST, 1991, 99 (02) : 358 - 362
  • [4] THE ACCURACY OF PORTABLE PEAK FLOW METERS
    MILLER, MR
    DICKINSON, SA
    HITCHINGS, DJ
    [J]. THORAX, 1992, 47 (11) : 904 - 909
  • [5] ACCURACY AND REPRODUCIBILITY OF THE ASSESS PEAK FLOW METER
    VANSCHAYCK, CP
    DOMPELING, E
    VANWEEL, C
    FOLGERING, H
    VANDENHOOGEN, HJM
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1990, 3 (03) : 338 - 341
  • [6] ACCURACY AND REPRODUCIBILITY OF THE ASSESS PEAK FLOW METER
    IMBRUCE, RP
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1991, 4 (05) : 632 - 632
  • [7] ACCURACY AND REPRODUCIBILITY OF THE ASSESS PEAK FLOW METER - REPLY
    VANSCHAYCK, CP
    DOMPLING, E
    FOLGERING, H
    VANWEEL, C
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1991, 4 (05) : 633 - 633
  • [8] Accuracy of portable devices in measuring peak cough flow
    Kulnik, Stefan Tino
    MacBean, Victoria
    Birring, Surinder Singh
    Moxham, John
    Rafferty, Gerrard Francis
    Kalra, Lalit
    [J]. PHYSIOLOGICAL MEASUREMENT, 2015, 36 (02) : 243 - 257
  • [9] Interinstrument variability of three portable peak flow meters
    Schmekel, B
    [J]. CHEST, 1996, 109 (06) : 1668 - 1668
  • [10] The portable laboratory:: an evaluation of the accuracy and reproducibility of i-STAT©
    Bingham, D
    Kendall, J
    Clancy, M
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 : 66 - 71