Accuracy of portable devices in measuring peak cough flow

被引:25
|
作者
Kulnik, Stefan Tino [1 ]
MacBean, Victoria [2 ]
Birring, Surinder Singh [2 ]
Moxham, John [2 ]
Rafferty, Gerrard Francis [2 ]
Kalra, Lalit [1 ]
机构
[1] Kings Coll London, Inst Psychiat Psychol & Neurosci, Dept Clin Neurosci, Stroke Res Team, London SE5 8AF, England
[2] Kings Coll Hosp London, Kings Coll London, Div Asthma Allergy & Lung Biol, Sch Med,Dept Resp Med & Allergy,Chest Unit, London SE5 9RS, England
基金
美国国家卫生研究院;
关键词
peak cough flow; accuracy; peak flow meter; spirometer; pneumotachograph; EXPIRATORY FLOW; MUSCLE STRENGTH; VOLUME; PREDICTORS; EFFICIENCY; MANEUVERS; CAPACITY;
D O I
10.1088/0967-3334/36/2/243
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Peak cough flow (PCF) measurements can be used as indicators of cough effectiveness. Portable peak flow meters and spirometers have been used to measure PCF, but little is known about their accuracy compared to pneumotachograph systems. The aim of this study was to compare the accuracy of four portable devices (Mini-Wright and Assess peak flow meters, SpiroUSB and Microlab spirometers) in measuring PCF with a calibrated laboratory based pneumotachograph system. Twenty healthy volunteers (mean (SD) age 45 (16) years) coughed through a pneumotachograph connected in series with each portable device in turn, and the differences in PCF readings were analysed. In addition, mechanically generated flow waves of constant peak flow were delivered through each device both independently and when connected in series with the pneumotachograph. Agreement between PCF readings obtained with the pneumotachograph and the portable devices was poor. Peak flow readings were on average lower by approximately 50 L min(-1) when measured using the portable devices; 95% limits of agreement spanned approximately 150 L min(-1). The findings highlight the potential for inaccuracy when using portable devices for the measurement of PCF. Depending on the measurement instrument used, absolute values of PCF reported in the literature may not be directly comparable.
引用
收藏
页码:243 / 257
页数:15
相关论文
共 50 条
  • [1] MEASURING PEAK COUGH FLOW: AN EVALUATION OF DEVICES AND INTERFACES
    Sheers, N.
    Saravanan, K.
    Retica, S.
    Howard, M.
    Berlowitz, D.
    [J]. RESPIROLOGY, 2021, 26 : 27 - 27
  • [2] MEASURING PEAK COUGH FLOW: AN EVALUATION OF DEVICES AND INTERFACES
    Sheers, N.
    Saravanan, K.
    Retica, S.
    Howard, M.
    Berlowitz, D.
    [J]. RESPIROLOGY, 2020, 25 : 217 - 217
  • [3] THE ACCURACY OF PORTABLE PEAK FLOW METERS
    MILLER, MR
    DICKINSON, SA
    HITCHINGS, DJ
    [J]. THORAX, 1992, 47 (11) : 904 - 909
  • [4] Comparison of peak cough flows measured by pneumotachograph and a portable peak flow meter
    Sancho, J
    Servera, E
    Díaz, J
    Marín, J
    [J]. AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 2004, 83 (08) : 608 - 612
  • [5] Peak cough flow measurement with a pneumotacograph and a portable peak flow meter inpatients with neuromuscular diseases
    Rodrigues, P. R.
    Brito, P. U.
    Fernandes, L.
    Rodrigues, C.
    Reis, A.
    MoitaCentro, J.
    [J]. REVISTA PORTUGUESA DE PNEUMOLOGIA, 2017, 23 (01) : 39 - 40
  • [6] MODELING THE ACCURACY AND PRECISION OF PORTABLE PEAK EXPIRATORY FLOW METERS
    BURTON, P
    SLY, P
    WILLET, K
    CAHILL, P
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 147 (04): : A267 - A267
  • [7] Accuracy in the Assessment of Cough Peak Flow: Good Progress for a "Work in Progress"
    Kulnik, Stefan Tino
    Lewko, Agnieszka
    MacBean, Victoria
    Spinou, Arietta
    [J]. RESPIRATORY CARE, 2020, 65 (01) : 133 - 134
  • [8] ACCURACY OF PEAK FLOW METERS - DONT INTERCHANGE DEVICES
    MCKENZIE, B
    [J]. BRITISH MEDICAL JOURNAL, 1994, 308 (6933): : 917 - 917
  • [9] CALIBRATION OF PEAK EXPIRATORY FLOW (PEF) MEASURING DEVICES
    MILLEDGE, JS
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1991, 4 (09) : 1152 - 1152
  • [10] COUGH PEAK FLOW RATE
    LEINER, GC
    ABRAMOWITZ, S
    SMALL, MJ
    STENBY, VB
    [J]. AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1966, 251 (02): : 211 - +