A COMPARISON OF TRANSCUTANEOUS, END-TIDAL AND ARTERIAL MEASUREMENTS OF CARBON-DIOXIDE DURING GENERAL-ANESTHESIA

被引:44
|
作者
REID, CW
MARTINEAU, RJ
MILLER, DR
HULL, KA
BAINES, J
SULLIVAN, PJ
机构
[1] OTTAWA GEN HOSP, DEPT ANAESTHESIA, 501 SMYTH RD, OTTAWA K1H 8L6, ONTARIO, CANADA
[2] OTTAWA GEN HOSP, DEPT BIOMED ENGN, OTTAWA K1H 8L6, ONTARIO, CANADA
[3] UNIV OTTAWA, OTTAWA K1N 6N5, ONTARIO, CANADA
关键词
CARBON DIOXIDE; MONITORING; TENSION; ARTERIAL; CUTANEOUS; END-TIDAL; GRADIENTS; VENTILATION; CARBON DIOXIDE TENSION;
D O I
10.1007/BF03008669
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A randomized, prospective study was performed to evaluate the accuracy of a new transcutaneous carbon dioxide (CO2) monitor (Fastrac(TM)) during general anaesthesia. Twenty-two adult patients undergoing elective surgery were subjected to three different levels of minute ventilation by varying their respiratory rates in a randomized cross-over design. Simultaneous measurements of transcutaneous CO2 (PTCCO2) and arterial CO2 (PaCO2) were obtained at three levels of minute ventilation (low, medium and high). End-tidal CO2 (PETCO2) values were also recorded from a mass spectrometer (SARA(TM)) at each time period. A total of 66 data sets with PaCO2 ranging from 28-62 mmHg were analyzed. The PTCCO2 values demonstrated a high degree of correlation with PaCO2 over the range of minute ventilation (y = 0.904x + 6.36, r = 0.92, P < 0.001). The PETCO2 measurement also demonstrated a generally good correlation with PaCO2 (y = 0.62x + 9.21, r = 0.89, and P < 0.01). However, the PETCO2-PaCO2 gradients (mean 7.0 +/- 3.1 mmHg) were greater than the PTCCO2-PaCO2 gradients (mean 2.3 +/- 2.4 mmHg) at all three levels of minute ventilation (P < 0.05). These differences were greatest when PaCO2 was in the high range (48-60 mmHg). We conclude that the new Fastrac(TM) CO2 monitor is accurate for monitoring carbon dioxide levels during general anaesthesia. The new transcutaneous devices provide an effective method for non-invasive monitoring of CO2 in situations where continuous, precise control of CO2 levels is desired.
引用
收藏
页码:31 / 36
页数:6
相关论文
共 50 条
  • [21] Low levels of end-tidal carbon dioxide during general anesthesia and postoperative mortality
    Solhpour, Ali
    CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2022, 69 (03): : 389 - 390
  • [22] Low levels of end-tidal carbon dioxide during general anesthesia and postoperative mortality
    Ali Solhpour
    Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2022, 69 : 389 - 390
  • [23] EFFECT OF PEEP ON THE ARTERIAL MINUS END-TIDAL CARBON-DIOXIDE GRADIENT
    BLANCH, L
    FERNANDEZ, R
    BENITO, S
    MANCEBO, J
    NET, A
    CHEST, 1987, 92 (03) : 451 - 454
  • [24] END-TIDAL CARBON-DIOXIDE CONCENTRATION DURING CARDIOPULMONARY RESUSCITATION
    KOESTER, DR
    NEW ENGLAND JOURNAL OF MEDICINE, 1988, 319 (09): : 579 - 579
  • [25] END-TIDAL CARBON-DIOXIDE CONCENTRATION DURING CARDIOPULMONARY RESUSCITATION
    FALK, JL
    RACKOW, EC
    WEIL, MH
    NEW ENGLAND JOURNAL OF MEDICINE, 1988, 318 (10): : 607 - 611
  • [26] An observational study of end-tidal carbon dioxide trends in general anesthesia
    Akkermans, Annemarie
    van Waes, Judith A. R.
    Thompson, Aleda
    Shanks, Amy
    Peelen, Linda M.
    Aziz, Michael F.
    Biggs, Daniel A.
    Paganelli, William C.
    Wanderer, Jonathan P.
    Helsten, Daniel L.
    Kheterpal, Sachin
    van Klei, Wilton A.
    Saager, Leif
    CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2019, 66 (02): : 149 - 160
  • [27] COLORIMETRIC END-TIDAL CARBON-DIOXIDE DETECTOR
    BHENDE, MS
    PEDIATRIC EMERGENCY CARE, 1995, 11 (01) : 58 - 61
  • [28] THE FEF END-TIDAL CARBON-DIOXIDE DETECTOR
    STRUNIN, L
    WILLIAMS, T
    ANESTHESIOLOGY, 1989, 71 (04) : 621 - 622
  • [29] A SPURIOUSLY LOW END-TIDAL CARBON-DIOXIDE
    HURLEY, MR
    PAULL, JD
    ANAESTHESIA AND INTENSIVE CARE, 1991, 19 (04) : 615 - 616
  • [30] IS CORRELATION OF ARTERIAL AND END-TIDAL CARBON-DIOXIDE PRESSURE DURING INHALATION ANESTHESIA BETTER WITH LARYNGEAL MASK AS COMPARED TO FACE MASK
    IVENS, D
    VERBORGH, C
    BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 : 9 - 9