A COMPARISON OF ONE VOLT JOSEPHSON-JUNCTION ARRAY VOLTAGE STANDARDS DRIVEN BY A COMMON MICROWAVE SOURCE

被引:1
|
作者
REYMANN, D [1 ]
LOHIVE, JP [1 ]
GENEVES, G [1 ]
机构
[1] BNM,LCIE,PRIMAIRE ELECT MAGNETISME LAB,F-92260 FONTENAY ROSES,FRANCE
关键词
D O I
10.1088/0026-1394/31/1/007
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have carried out a direct comparison between the BIPM and the BNM/LCIE Josephson junction arrays driving them with a single microwave source. The mean difference observed at the 1,018 V level is, 1,2 parts in 10(10) with a total uncertainty. below 1,7 parts in 10(10). By removing uncertainties due to frequency instability, the technique allows closer examination of the other limiting factors in these measurements. In this case we studied the resolution of different detectors.
引用
收藏
页码:35 / 37
页数:3
相关论文
共 41 条
  • [31] IMPLEMENTATION OF JOSEPHSON TUNNEL JUNCTION SERIES ARRAY STRIPLINE WITH HIGH CHARACTERISTIC IMPEDANCE FOR QUANTUM VOLTAGE STANDARDS
    GUTMANN, P
    VOLLMER, E
    KOHLMANN, J
    MULLER, F
    WEIMANN, T
    NIEMEYER, J
    [J]. ELECTRONICS LETTERS, 1992, 28 (15) : 1422 - 1423
  • [32] New microwave circuits for programmable voltage standards using high-temperature Josephson junction arrays
    Klushin, AM
    Weber, C
    Kohlstedt, H
    Starodubrovskii, RK
    Lauer, A
    Wolff, I
    Semerad, R
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 4158 - 4161
  • [33] Indirect comparison of 10 V Josephson array voltage standards between the UME and BNM-LCIE
    Selçik, S
    Lo-Hive, JP
    Yilmaz, O
    Genevès, G
    [J]. METROLOGIA, 2001, 38 (05) : 477 - 480
  • [34] PTB'S STATUS OF AC VOLTAGE METROLGY BY MICROWAVE- AND PULSETRAIN-DRIVEN JOSEPHSON JUNCTION ARRAYS
    Niemeyer, J.
    Kieler, O.
    Mueller, F.
    Kohlmann, J.
    Behr, R.
    Palafox, L.
    Schleussner, D.
    Beug, M. F.
    Schurr, J.
    [J]. 2013 INTERNATIONAL KHARKOV SYMPOSIUM ON PHYSICS AND ENGINEERING OF MICROWAVES, MILLIMETER AND SUBMILLIMETER WAVES (MSMW), 2013, : 652 - 654
  • [35] DEVELOPMENT OF A HIGH-QUALITY NB/AL-OXIDE/NB JOSEPHSON-JUNCTION ARRAY (0.1-V MODEL) FOR A VOLTAGE STANDARD
    PARK, SI
    KIM, KT
    LEE, YH
    LEE, RD
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 1993, 26 (04) : 369 - 377
  • [37] Evaluation of the Zero-Compensation Method Based on the Comparison of Two Pulse-Driven Josephson Voltage Standards
    Kraus, M.
    Behr, R.
    Kieler, O.
    Bauer, S.
    Palafox, L.
    Herick, J.
    [J]. 2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [38] HYSTERESIS LOOP IN CURRENT-VOLTAGE CURVE FOR BAPB0.7BI0.3O3 JOSEPHSON JUNCTION ARRAY IN A MICROWAVE FIELD
    ITO, M
    ENOMOTO, Y
    SUZUKI, M
    MURAKAMI, T
    INAMURA, T
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1982, 21 (06): : L375 - L376
  • [39] Bilateral direct comparison of Josephson array voltage standards of the PTB (Germany) and PEL (Croatia): Draft B report for EUROMET.EM-S28
    Behr, Ralf
    Ilic, Damir
    Sala, Alan
    [J]. METROLOGIA, 2008, 45
  • [40] Proximity-Driven Source of Highly Spin-Polarized ac Current on the Basis of Superconductor/Weak Ferromagnet/Superconductor Voltage-Biased Josephson Junction
    Bobkov, A. M.
    Bobkova, I. V.
    [J]. JETP LETTERS, 2008, 88 (12) : 807 - 813