TEST OF LONG SCINTILLATION-COUNTERS FOR SUPERCOLLIDER DETECTORS

被引:12
|
作者
BELLAMY, EH
BELLETTINI, G
BUDAGOV, J
CERVELLI, F
CHIRIKOVZORIN, I
KOVTUN, V
INCAGLI, M
LUCCHESI, D
PAGLIARONE, C
PUKHOV, O
SEMINOZHENKO, V
SENCHISHIN, V
TOKAR, S
VEREZUB, N
ZALJUBOVSKY, I
ZETTI, F
机构
[1] UNIV PISA,DIPARTIMENTO FIS,I-56100 PISA,ITALY
[2] IST NAZL FIS NUCL,SEZ PISA,PISA,ITALY
[3] SUPERCONDUCTING SUPER COLLIDER LAB,DALLAS,TX
[4] DUBNA JOINT NUCL RES INST,DUBNA,RUSSIA
[5] KHARKOV AM GORKII STATE UNIV,KHARKOV,UKRAINE
[6] STC INST MONOCRYSTAL SA UKRAINE,KHARKOV,UKRAINE
[7] COMENIUS UNIV BRATISLAVA,CS-80100 BRATISLAVA,SLOVAKIA
[8] POLYTECH INST,KHARKOV,UKRAINE
关键词
D O I
10.1016/0168-9002(94)90228-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A number of long scintillator counters employing UPS 923A and NE 114 plastic scintillators seen by a single photomultiplier were tested using cosmic muons. These counters have suitable dimensions to be used in large arrays in a proton supercollider experiment. The efficiency for m.i.p. was excellent in all cases. The yield of UPS 923A was found to be by about 25% higher than NE 114, while the attenuation length was approximately the same.
引用
收藏
页码:484 / 488
页数:5
相关论文
共 50 条
  • [1] TIMING WITH SCINTILLATION-COUNTERS
    ELWAHAB, MA
    BATTRAWI, MH
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) : 625 - 629
  • [2] THE TEST MEASUREMENTS OF TOF IN SMALL AND LARGE SCINTILLATION-COUNTERS
    IWATA, T
    DOHMAE, K
    FUKUI, S
    GAO, CY
    HIROSE, I
    HORIKAWA, N
    NAKANISHI, T
    SAKANAKA, S
    TOYAMA, T
    KITAGAKI, T
    OHTANI, Y
    SAGAWA, H
    YAMAGUCHI, A
    MUTOH, M
    YAMAMOTO, R
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 244 (03): : 434 - 439
  • [3] AN ANALOG MEAN-TIMER FOR LONG SCINTILLATION-COUNTERS
    CAVESTRO, A
    GIBIN, D
    GUGLIELMI, A
    MATTIOLI, F
    MEZZETTO, M
    SCHIAVON, PG
    PUGLIERIN, G
    VASCON, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 305 (02): : 488 - 491
  • [4] SCINTILLATION EFFICIENCY OF GAS PROPORTIONAL SCINTILLATION-COUNTERS
    ALEGRIAFEIO, M
    POLICARPO, A
    CHARPAK, G
    SAULI, F
    NUCLEAR INSTRUMENTS & METHODS, 1980, 176 (03): : 473 - 476
  • [5] TEST OF 2-MM-THICK TOF SCINTILLATION-COUNTERS
    FUJII, Y
    OSHIMA, T
    SHIRAI, J
    SUGIMOTO, Y
    YASU, Y
    HARADA, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 366 (2-3): : 282 - 291
  • [6] HIGHLY EFFICIENT SCINTILLATION-COUNTERS
    MOUNT, R
    CAVAGLIA, P
    STOCKHAUSEN, W
    WAHLEN, H
    NUCLEAR INSTRUMENTS & METHODS, 1979, 160 (01): : 23 - 27
  • [7] TIMING PROPERTIES OF LONG SCINTILLATION-COUNTERS BASED ON SCINTILLATING FIBERS
    KUHLEN, M
    MOSZYNSKI, M
    STROYNOWSKI, R
    WICKLUND, E
    MILLIKEN, B
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 301 (02): : 223 - 229
  • [8] ANSI STANDARDS FOR LIQUID SCINTILLATION-COUNTERS
    FERRIS, R
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1980, 31 (01): : 10 - 10
  • [9] INVESTIGATION OF AMPLITUDE CHARACTERISTICS OF SCINTILLATION-COUNTERS
    GLAVANAKOV, IV
    STIBUNOV, VN
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1978, 21 (01) : 32 - 35
  • [10] COMPARISON OF REFLECTORS USED IN SCINTILLATION-COUNTERS
    SALOMON, M