EARLY EXPERIENCE OF A COMMERCIAL SCANNED FOCUSED ULTRASOUND HYPERTHERMIA SYSTEM

被引:21
|
作者
HAND, JW
VERNON, CC
PRIOR, MV
机构
[1] Hyperthermia Clinic, MRC Cyclotron Unit, Hammersmith Hospital, London, W12 OHS, Du Cane Road
关键词
ULTRASOUND; CLINICAL HYPERTHERMIA; DEEP HYPERTHERMIA; FOCUSED BEAMS;
D O I
10.3109/02656739209037995
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Preliminary experiences of the Sonotherm 6500 scanned focused ultrasound (SFUS) system in the hyperthermic treatment of bulky tumours in breast, superficial sites and within the pelvis in 22 patients are reported. Tumour volumes ranged from 235 to 603 cm3 (breast). 105 to 209 cm3 (superficial sites) and 24 to 905 cm3 (pelvis). Temperature distributions in 58 evaluable treatments were analysed in terms of temporal peak and time-averaged temperatures (highest and lowest temperatures, percentage of sensors exceeding index temperatures, etc.) achieved within scanned volumes. Mean number of sensors implanted into tumour was 14 in breast and superficial tumours and approximately 11 in pelvic tumours. Mean time-averaged maximum and minimum temperatures for the best treatments (i.e. for each patient, that with the highest percentage of sensors recording time averaged temperatures greater-than-or-equal-to 42-degrees-C) administered to patients with tumours in the breast and superficial sites were 44.6 +/- 1.7-degrees-C and 39.7 +/- 1.1-degrees-C, respectively and the mean number of sensors exceeding 42-degrees-C was 58 +/- 19%. In the case of pelvic tumours these figures were 41.6 +/- 0.9-degrees-C, 40.0 +/- 0.6-degrees-C and 8 +/- 12%, respectively. Patient tolerance to treatments was, in general, good. Areas where technical improvement of the system is appropriate and further research and development are required are identified; these should lead to a better realization of the potential of the SFUS technique, particularly for pelvic tumours.
引用
收藏
页码:587 / 607
页数:21
相关论文
共 50 条
  • [1] FOCUSED, SCANNED ULTRASOUND FOR LOCAL HYPERTHERMIA
    HYNYNEN, K
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (10) : 877 - 877
  • [3] DOSE UNIFORMITY IN SCANNED FOCUSED ULTRASOUND HYPERTHERMIA
    LAGENDIJK, JJW
    CREZEE, J
    HAND, JW
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1994, 10 (06) : 775 - 784
  • [4] Control of scanned, focused ultrasound hyperthermia treatments
    Lin, Win-Li
    Wang, C.Y.
    Shieh, M.J.
    Biomedical Engineering - Applications, Basis and Communications, 1992, 4 (06): : 618 - 625
  • [5] MULTIPOINT FEEDBACK-CONTROL SYSTEM FOR SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA
    JOHNSON, C
    KRESS, R
    ROEMER, R
    HYNYNEN, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (06): : 781 - 786
  • [6] NONLINEAR ABSORPTION DURING SCANNED FOCUSED ULTRASOUND HYPERTHERMIA
    HYNYNEN, K
    ROEMER, RB
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1986, 33 (01) : 107 - 107
  • [7] OPTIMIZATION OF TEMPERATURE DISTRIBUTIONS IN SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA
    LIN, WL
    ROEMER, RB
    MOROS, EG
    HYNYNEN, K
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (01) : 61 - 78
  • [8] SCANNED FOCUSED ULTRASOUND HYPERTHERMIA - INITIAL CLINICAL-RESULTS
    SHIMM, DS
    HYNYNEN, KH
    ANHALT, DP
    ROEMER, RB
    CASSADY, JR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 15 (05): : 1203 - 1208
  • [9] HYPERTHERMIA TREATMENT OF BRAIN-TUMORS USING SCANNED, FOCUSED ULTRASOUND
    ROEMER, RB
    HYNYNEN, K
    TOBIAS, J
    FLEISHER, A
    GUTHKELCH, AN
    TOBIAS, J
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (06) : 588 - 588
  • [10] DEVELOPMENT OF SCANNED FOCUSED ULTRASOUND HYPERTHERMIA - CLINICAL-RESPONSE EVALUATION
    HARARI, PM
    HYNYNEN, KH
    ROEMER, RB
    ANHALT, DP
    SHIMM, DS
    STEA, B
    CASSADY, JR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03): : 831 - 840