ANALYSIS OF BUMETANIDE IN HUMAN URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION AND GAS-CHROMATOGRAPHY MASS-SPECTROMETRY

被引:15
|
作者
GRADEEN, CY [1 ]
BILLAY, DM [1 ]
CHAN, SC [1 ]
机构
[1] FOOTHILLS PROV GEN HOSP,DEPT LAB MED,1403 29TH ST NW,CALGARY T2N 2T9,ALBERTA,CANADA
关键词
D O I
10.1093/jat/14.2.123
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Bumetanide is a potent diuretic. Its concentration in urine after therapeutic doses is in the low ng/mL range. A procedure using reversed-phase HPLC with fluorescence detection is described. Using this procedure, the authors have been able to measure bumetanide in a urine sample at 10 ng/mL with a signal-to-noise ratio of 7. Therefore, the detection limit is lower than this concentration. Confirmation is by a GC/MS procedure after methylating the diuretic. Derivatization is by extractive alkylation with tetrabutylammonium hydrogen sulfate as the phase transfer reagent and iodomethane as the methylating agent. © 1990 Oxford University Press.
引用
收藏
页码:123 / 126
页数:4
相关论文
共 50 条