PINEAL AND PLASMA MELATONIN AS DETERMINED BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION

被引:61
|
作者
VIEIRA, R
MIGUEZ, J
LEMA, M
ALDEGUNDE, M
机构
[1] Laboratorio de Fisioloxía Animal, Departamento de Fisioloxía, Facultade de Bioloxía
关键词
D O I
10.1016/0003-2697(92)90439-E
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid and sensitive method for the routine quantitative determination of melatonin in pineal and plasma is described. The assay used reversed-phase high-performance liquid chromatography (RP-HPLC) separation combined with either amperometric (system A) or coulometric (system B) detection. The method gave satisfactory reproducibility and accuracy, and detection limits for melatonin were as low as 8.5 pg (system A) and 1 pg (system B). This high sensitivity, together with the short analysis time (less than 10 min), and the simplicity of sample procedure make the present RP-HPLC method suitable for a wide range of studies concerning melatonin measurements. Melatonin values obtained in this study from both rat pineal and human plasma agree with those reported previously, and clearly determined a circadian pattern. © 1992.