Dosimetrical evaluation of Leksell Gamma Knife 4C radiosurgery unit

被引:0
|
作者
Sajeev, Thomas [1 ,2 ]
Mustafa, Mohamed M. [2 ]
Supe, Sanjay S. [3 ]
机构
[1] Deemed Univ, Natl Inst Mental Hlth & Neuro Sci, Dept Neurosurg, Bangalore, Karnataka, India
[2] Univ Calicut, Dept Phys, Kozhikode, Kerala, India
[3] Kidwai Mem Inst Oncol, Dept Radiat Phys, Banaglore, India
关键词
Gamma knife; radiosurgery; dosimetry; Leksell; stereotactic; accuracy;
D O I
10.2478/v10013-011-0011-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A number of experiments was performed using standard protocols, in order to evaluate the dosimetric accuracy of Leksell Gamma Knife 4C unit. Verification of the beam alignment has been performed for all collimators using solid plastic head phantom and Gafchromic TM type MD-55 films. The study showed a good agreement of Leksell Gammaplan calculated dose profiles with experimentally determined profiles in all three axes. Isocentric accuracy is verified using a specially machined cylindrical aluminium film holder tool made with very narrow geometric tolerances aligned between trunnions of 4 mm collimator. Considering all uncertainties in all three dimensions, the estimated accuracy of the unit was 0.1 mm. Dose rate at the centre point of the unit has been determined according to the IAEA, TRS-398 protocol, using Unidose-E (PTW-Freiburg, Germany) with a 0.125 cc ion chamber, over a period of 6 years. The study showed that the Leksell Gamma Knife 4C unit is excellent radiosurgical equipment with high accuracy and precision, which makes it possible to deliver larger doses of radiation, within the limits defined by national and international guidelines, applicable for stereotactic radiosurgery procedures.
引用
收藏
页码:105 / 120
页数:16
相关论文
共 50 条
  • [31] Monte Carlo calculated output factors of a Leksell Gamma Knife unit
    Cheung, JYC
    Yu, KN
    Ho, RTK
    Yu, CP
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12): : N247 - N249
  • [32] Stereotactic radiosurgery of essential trigeminal neuralgia using Leksell Gamma Knife model C with automatic positioning system
    Hayashi, Motohiro
    Chernov, Mikhail
    Tamura, Noriko
    Taira, Takaomi
    Izawa, Masahiro
    Yomo, Shoji
    Nagai, Mariko
    Chang, Cheng-Siu
    Ivanov, Pavel
    Tamura, Manabu
    Muragaki, Yoshihiro
    Okada, Yoshikazu
    Iseki, Hiroshi
    Takakura, Kintomo
    NEUROSURGICAL REVIEW, 2011, 34 (04) : 497 - 508
  • [33] ANOMALOUS PRIMARY RADIATION FROM THE LEKSELL GAMMA-KNIFE UNIT
    SHEARER, DR
    ROVENTINE, WA
    COY, SR
    HEALTH PHYSICS, 1994, 67 (06): : 657 - 660
  • [34] An investigation of eye lens dose of stereotactic radiosurgery for trigeminal neuralgia using Leksell Gamma Knife model C
    Liang, Cheng-Loong
    Ho, Meng-Wei
    Lu, Kang
    Tsai, Yu-Duan
    Liliang, Po-Chou
    Wang, Kuo-Wei
    Chen, Han-Jung
    JOURNAL OF NEUROSURGERY, 2006, 105 : 112 - 116
  • [35] Optically guided Linac radiosurgery with a leksell head frame as an adjunct to Gamma Knife™ treatment
    Ekstrand, Kenneth E.
    Hinson, William H.
    Kearns, William
    Bourland, J. Daniel
    deGuzman, Allan F.
    McMullen, Kevin P.
    Stieber, Volker W.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2007, 6 (02) : 123 - 126
  • [36] Arteriovenous malformations after leksell gamma knife radiosurgery:: Rate of obliteration and complications -: Comments
    Pollock, Bruce E.
    Lunsford, L. Dade
    Tamura, Manabu
    Regis, Jean
    Duckworth, Edward
    Batjer, H. Hunt
    Sakamoto, Gordon T.
    Steinberg, Gary K.
    NEUROSURGERY, 2007, 60 (06) : 1015 - 1016
  • [37] SRS Stereotactic RadioSurgery using Elekta Leksell Gamma Knife Perfexion (GK) device
    Kawa-Iwanicka, A.
    Nawrocki, S.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : II - II
  • [38] Impact of the skull contour definition on Leksell Gamma Knife® Icon™ radiosurgery treatment planning
    Henri-Arthur Leroy
    Constantin Tuleasca
    Michele Zeverino
    Elodie Drumez
    Nicolas Reyns
    Marc Levivier
    Acta Neurochirurgica, 2020, 162 : 2203 - 2210
  • [39] A Monte Carlo model and its commissioning for the Leksell Gamma Knife Perfexion radiosurgery system
    Yuan, Jiankui
    Machtay, Mitchell
    MEDICAL PHYSICS, 2017, 44 (09) : 4910 - 4918
  • [40] Leksell Gamma Knife radiosurgery of the jugulotympanic glomus tumor: long-term results
    Liscak, Roman
    Urgosik, Dusan
    Chytka, Tomas
    Simonova, Gabriela
    Novotny, Josef, Jr.
    Vymazal, Josef
    Guseynova, Khumar
    Vladyka, Vilibald
    JOURNAL OF NEUROSURGERY, 2014, 121 : 198 - 202