DEVELOPMENT OF SEMIPORTABLE DC-SQUID MAGNETOMETER

被引:5
|
作者
KANDORI, A [1 ]
UEDA, M [1 ]
OGATA, H [1 ]
KADO, H [1 ]
机构
[1] ELECTROTECH LAB, TSUKUBA, IBARAKI 305, JAPAN
关键词
D O I
10.1109/77.403092
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A semi-portable DC-SQUID magnetometer with a liquid helium continuous how cryostat has been developed. The cryostat is 340 mm long and 50 mm in outer diameter. The pick-up coil is a co-axial, first order, differential type, 20 mm in diameter and 30 mm in baseline. The distance between the pick-up coil and the outer surface of the cryostat is 5 mm. It is easy to set this cryostat on the chest wall or on a human head. The cool down time is about 20 minutes, which is much shorter than that for a conventional storage Dewar. The white noise level, measured in a magnetically shielded room, is 8.7 fT/root Hz at 30 Hz. Noise at 1 Hz is about 30 fT/root Hz. Noise from temperature fluctuations (< 9 mK) and drift (4.22 K similar to 4.55 K) is negligible. The consumption rate of liquid helium is 10 l/h. In this paper, we show the measurement of a magnetocardiogram in a magnetically shielded room.
引用
收藏
页码:2474 / 2477
页数:4
相关论文
共 50 条
  • [1] DC-SQUID PREAMPLIFIER FOR DC-SQUID MAGNETOMETER
    KOSHELETS, VP
    MATLASHOV, AN
    SERPUCHENKO, IL
    FILIPPENKO, LV
    ZHURAVLEV, YE
    IEEE TRANSACTIONS ON MAGNETICS, 1989, 25 (02) : 1182 - 1185
  • [2] INTEGRATED DC-SQUID MAGNETOMETER
    NAKANISHI, M
    KOYANAGI, M
    KOSAKA, S
    SHOJI, A
    AOYAGI, M
    SHINOKI, F
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1987, 26 (07): : 1050 - 1055
  • [3] Development of semi-portable DC-SQUID magnetometer
    Kandori, A.
    Ueda, M.
    Ogata, H.
    Kado, H.
    IEEE Transactions on Applied Superconductivity, 1995, 5 (2 pt 3): : 2474 - 2477
  • [4] INTEGRATED DC-SQUID MAGNETOMETER.
    Nakanishi, Masakazu
    Koyanagi, Masao
    Kosaka, Shin
    Shoji, Akira
    Aoyagi, Masahiro
    Shinoki, Fujitoshi
    Japanese Journal of Applied Physics, Part 1: Regular Papers & Short Notes, 1987, 26 (07): : 1050 - 1055
  • [5] A DOUBLE DC-SQUID BASED MAGNETOMETER
    MASLENNIKOV, YV
    BELJAEV, AV
    SNIGIREV, OV
    KAPLUNENKO, OV
    MEZZENA, R
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1995, 5 (02) : 3241 - 3243
  • [6] Dynamics and Noise in dc-SQUID Magnetometer Arrays
    Aven, John L.
    Palacios, Antonio
    Longhini, Patrick
    In, Visarath
    Bulsara, Adi
    APPLICATIONS OF NONLINEAR DYNAMICS-MODEL AND DESIGN OF COMPLEX SYSTEMS, 2009, : 381 - +
  • [7] Development of mice biomagnetic measurement system using dc-SQUID magnetometer
    Ono, Y
    Ishiyama, A
    Kasai, N
    Odawara, A
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2005, 15 (02) : 640 - 643
  • [8] OPTIMIZATION OF A DC-SQUID MAGNETOMETER TO MINIMIZE THE FIELD RESOLUTION
    SUZUKI, K
    OKABE, Y
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1995, 5 (02) : 2172 - 2175
  • [9] A 37-CHANNEL DC-SQUID MAGNETOMETER SYSTEM
    DRUNG, D
    ZIMMERMANN, R
    CANTOR, R
    ERNE, SN
    KOCH, H
    MATTHIES, KP
    PETERS, M
    SCHEER, HJ
    STOLLFUSS, D
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1991, 12 : 21 - 29
  • [10] INTEGRATED DC-SQUID MAGNETOMETER WITH SIMPLIFIED READ-OUT
    CANTOR, R
    DRUNG, D
    PETERS, M
    SCHEER, HJ
    KOCH, H
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1990, 3 (02): : 108 - 112