Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures

被引:10
|
作者
Sukupova, Lucie [1 ]
Hlavacek, Ondrej [2 ]
Vedlich, Daniel [2 ]
机构
[1] Inst Clin & Expt Med, Dept Director, Videnska 1958-9, Prague 14021 4, Czech Republic
[2] Inst Clin & Expt Med, Radiodiagnost & Intervent Radiol Dept, Videnska 1958-9, Prague 14021 4, Czech Republic
关键词
D O I
10.1155/2018/4287973
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters. Measurements were performed in different projections and with different positions of the ceiling-mounted shielding: without shielding; shielding closest to the patient; and shielding closest to the physician performing the procedure. The protection provided by the shielding was assessed for cardiology when the femoral access is used and for radiology when the physician performs the procedure in the abdominal area. The results show that the use of the ceiling-mounted shielding can decrease the dose from the scatter radiation by 95% at the position of the performing physician. In cardiology, the impact is more pronounced when the left oblique projection is used. In radiology, a large decrease was observed for right oblique projections, compared to cardiology. The ceiling-mounted shielding should be placed as close to the physician as possible. The idea of creating the largest radiation shadow by placing the radiation shielding as close to the patient as possible does not provide as effective radiation protection of the operator as it might be thought.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Evaluation of thyroid radiation dose during abdominal Computed Tomography procedures and dose reduction effectiveness of thyroid shielding
    Jeyasugiththan, J.
    Kumarihami, A. M. C.
    Satharasinghe, D.
    Mahakumara, P.
    Senanayaka, G.
    Jayakody, I.
    RADIOGRAPHY, 2022, 28 (03) : 704 - 710
  • [22] Scatter Radiation Measurement From a Digital Breast Tomosynthesis System and Its Impact On Shielding Consideration
    Yang, K.
    Li, X.
    Liu, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3819 - 3819
  • [23] Canadian Cardiovascular Society Position Statement on Radiation Exposure From Cardiac Imaging and Interventional Procedures
    Natarajan, Madhu K.
    Paul, Narinder
    Mercuri, Mathew
    Waller, Edward J.
    Leipsic, Jonathon
    Traboulsi, Mouhieddin
    Banijamali, Hamid S.
    Benson, Lee
    Sheth, Tej N.
    Simpson, Christopher S.
    Brydie, Allan
    Love, Michael P.
    Gallo, Richard
    CANADIAN JOURNAL OF CARDIOLOGY, 2013, 29 (11) : 1361 - 1368
  • [24] Reduced radiation dose in CT guided interventional procedures in the abdomen and pelvis: Does it impact on technical success?
    Lucey, BC
    Blake, MA
    Kalra, MK
    Maher, MM
    Saini, S
    Mueller, PR
    RADIOLOGY, 2002, 225 : 252 - 252
  • [25] Analysis of factors that affect radiation dose level during interventional cardiology procedures using logistic regression
    Harki, Edrees M.
    Rashid, Z. S.
    REVISTA MEXICANA DE FISICA, 2023, 69 (03)
  • [26] Development and evaluation of modified lead gloves to reduce hand radiation dose during interventional radiological procedures
    Park, Hyun-Jun
    Jo, Byungdu
    Lee, Seung-Jae
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2024, 56 (07) : 2781 - 2789
  • [27] A preliminary study on the radiation dose reduction during catheterization procedures using simple partial patient shielding
    Serati, Alireza
    Hashemi, Ashkan
    Sharifkashani, Babak
    Nourbakhsh, Sourmah
    Hashemi, Arash
    Emami, Sepideh
    Movahed, Mohammad Reza
    REVIEWS IN CARDIOVASCULAR MEDICINE, 2021, 22 (03) : 891 - 894
  • [28] Usefulness of an additional lead shielding device in reducing occupational radiation exposure during interventional endoscopic procedures An observational study
    Yamada, Reiko
    Saimyo, Yusuke
    Tanaka, Kyosuke
    Hattori, Aiji
    Umeda, Yuhei
    Kuroda, Naoki
    Tsuboi, Junya
    Hamada, Yasuhiko
    Takei, Yoshiyuki
    MEDICINE, 2020, 99 (34) : E21831
  • [29] CHARACTERIZATION AND EXPERIMENTAL MEASUREMENTS OF SCATTER DOSE AT CARDIOLOGIST'S EYES DURING PAEDIATRIC INTERVENTIONAL CARDIOLOGY PROCEDURES IN COSTA RICA
    Ubeda, Carlos
    Salazar, Luisa
    Retana, Viviana
    Santos, Freddy
    Salvador, Lourdes
    Saenz, Carlos
    Quesada, Cristhian
    Gavarrete, Manuel
    Picado, Marcela
    Arce, Luis
    RADIATION PROTECTION DOSIMETRY, 2017, 176 (04) : 450 - 455
  • [30] Development of a real-time scattered radiation display for staff dose reduction during fluoroscopic interventional procedures
    Troville, J.
    Kilian-Meneghin, J.
    Guo, C.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948