A position-sensitive ion detection system for trace analysis with magnetic sector mass spectrometers is described in detail, with particular application to high temperature mass spectrometry. The detection system consists of two stacked microchannel plates (Chevron assembly) backed by a resistive anode encoder and associated electronics. The range of masses simultaneously detectable is m to 1.2m. For election impact ionization of silver at an electron energy of 10.5 eV, the sensitivity is 1.6 x 10(-7) pa, and the mass resolution is 260 at mass 80 (valley 10% of the peak height definition). Additional applications for the detection system are discussed.
引用
收藏
页码:661 / 676
页数:16
相关论文
共 50 条
  • [41] LASER IONIZATION MASS-SPECTROMETRY IN INORGANIC TRACE ANALYSIS
    BECKER, JS
    DIETZE, HJ
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 344 (03): : 69 - 86
  • [42] TIME-RESOLVED MASS AND ENERGY ANALYSIS BY POSITION-SENSITIVE TIME-OF-FLIGHT DETECTION
    POLLARD, JE
    LICHTIN, DA
    JANSON, SW
    COHEN, RB
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (10): : 3171 - 3180
  • [43] DETECTION OF TRACE AMOUNTS OF ACTINIDES BY RESONANCE IONIZATION MASS-SPECTROMETRY
    RIMKE, H
    HERRMANN, G
    MUHLECK, C
    SATTELBERGER, P
    TRAUTMANN, N
    AMES, F
    KLUGE, HJ
    OTTEN, EW
    REHKLAU, D
    RUSTER, W
    INORGANICA CHIMICA ACTA, 1987, 140 (1-2) : 277 - 278
  • [44] TRACE DETECTION OF RADIOTOXIC ISOTOPES BY RESONANCE IONIZATION MASS-SPECTROMETRY
    WENDT, K
    PASSLER, G
    TRAUTMANN, N
    PHYSICA SCRIPTA, 1995, T58 : 104 - 108
  • [45] TIME-RESOLVED MASS AND ENERGY ANALYSIS BY POSITION-SENSITIVE TIME-OF-FLIGHT DETECTION
    POLLARD, JE
    LICHTIN, DA
    JANSON, SW
    COHEN, RB
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10): : 3134 - 3136
  • [46] OPTIMIZATION OF A COAXIAL CONTINUOUS-FLOW FAST ATOM BOMBARDMENT INTERFACE BETWEEN CAPILLARY LIQUID-CHROMATOGRAPHY AND MAGNETIC-SECTOR MASS-SPECTROMETRY FOR THE ANALYSIS OF BIOMOLECULES
    MOSELEY, MA
    DETERDING, LJ
    DEWIT, JSM
    TOMER, KB
    KENNEDY, RT
    BRAGG, N
    JORGENSON, JW
    ANALYTICAL CHEMISTRY, 1989, 61 (14) : 1577 - 1584
  • [47] EFFICIENCY OF BE-8 DETECTION USING POSITION-SENSITIVE DETECTORS
    MENCHACAROCHA, A
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1975, 22 (04) : 1997 - 2000
  • [48] Compact broadly-tunable UVOPO for trace detection using REMPI mass-spectrometry
    Barnes, RJ
    Margalith, E
    Oser, H
    Coggiola, M
    Young, S
    Crosley, D
    SOLID STATE LASERS XIII: TECHNOLOGY AND DEVICES, 2004, 5332 : 251 - 262
  • [49] IMPROVED DETECTION LIMITS FOR ELECTROSPRAY-IONIZATION ON A MAGNETIC-SECTOR MASS-SPECTROMETER BY USING AN ARRAY DETECTOR
    CODY, RB
    TAMURA, J
    FINCH, JW
    MUSSELMAN, BD
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (03) : 194 - 200
  • [50] MULTIPLE ION DETECTION SYSTEM WITH MINISCAN FACILITIES AND EXPANDED MASS RANGE FOR MAGNETIC-SECTOR MASS SPECTROMETERS
    ARTIGAS, F
    GELPI, E
    PRUDENCIO, M
    ALONSO, JA
    BAILLART, J
    ANALYTICAL CHEMISTRY, 1977, 49 (04) : 543 - 549