ON THE PERFORMANCE OF THE BERNER LOW-PRESSURE IMPACTOR

被引:209
|
作者
HILLAMO, RE [1 ]
KAUPPINEN, EI [1 ]
机构
[1] TECH RES CTR FINLAND,HEATING & VENTILATING LAB,SF-02150 ESPOO,FINLAND
基金
芬兰科学院;
关键词
D O I
10.1080/02786829108959469
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Particle collection characteristics of the compressible flow stages 1-6 of the Berner low-pressure impactor (BLPI, Hauke Aeras 25-4/0.015) have been studied. A new, real-time impactor stage efficiency measurement system has been developed. It is based on the use of two electrometers, one connected to the collection plate of the impactor, and the other to the backup filter. This method is used with singly charged, monodisperse dioctyl phthalate (DOP) particles to determine the shape of the collection efficiency curves of the BLPI. In addition, particle collection characteristics have been studied using fluorometric analysis. A new method for substrate film greasing is also introduced. Experimental cut diameters of stages 1-6, as determined with DOP particles using electrical measurement method, are 0.034, 0.066, 0.094, 0.17, 0.33, and 0.55-mu-m. They are clearly larger than corresponding values given by the manufacturer. The average of the square root of the Stokes number at 50% collection efficiency, based on experimental cut diameters and evaluated at stage conditions, is 0.465. For solid ammonium fluorescein particles the collection efficiency of high-velocity stages (1-6) was poor, although the substrates (Mylar film) were greased with Apiezon L vacuum grease. Using experimentally determined stage pressures and calculated cut diameters it was established that the compressible flow stages of different units of the same BLPI model may have significantly different cut diameters. The effect of temperature and pressure on cutoff diameters at atmospheric conditions near ground level is also discussed.
引用
收藏
页码:33 / 47
页数:15
相关论文
共 50 条
  • [21] SIMPLE NITRIC-ACID DISSOLUTION METHOD FOR ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRIC ANALYSIS OF ATMOSPHERIC AEROSOL SAMPLES COLLECTED BY A BERNER-TYPE LOW-PRESSURE IMPACTOR
    PAKKANEN, TA
    HILLAMO, RE
    MAENHAUT, W
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) : 79 - 84
  • [22] PERFORMANCE OF LOW-PRESSURE DROP PACKINGS
    BILLET, R
    CHEMICAL ENGINEERING COMMUNICATIONS, 1987, 54 (1-6) : 93 - 118
  • [23] PERFORMANCE TESTING OF A LOW-PRESSURE COMPRESSOR
    Sans, J.
    Desset, J.
    Dell'Era, G.
    Brouckaert, J. -F.
    Hiernaux, S.
    10TH EUROPEAN CONFERENCE ON TURBOMACHINERY: FLUID DYNAMICS AND THERMODYNAMICS, 2013,
  • [24] Characterization of a low-pressure exchanging virtual impactor for in situ analysis of atmospheric aerosols.
    Boulter, JE
    Middlebrook, AM
    Murphy, DM
    Birks, JW
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2000, 219 : U316 - U316
  • [26] Calibration of the Modified Electrical Low-Pressure Impactor (ELPI) for Use with Pressurized Pharmaceutical Aerosols
    Kotian, Reshma
    Peart, Joanne
    Bryner, Joan
    Byron, Peter R.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2009, 22 (01) : 55 - 65
  • [27] FLOW-ANALYSIS OF A LOW-PRESSURE IMPACTOR BY COMPRESSIBLE NAVIER-STOKES EQUATION
    KOGA, J
    NAMMO, A
    MATSUMOTO, S
    JOURNAL OF CHEMICAL ENGINEERING OF JAPAN, 1995, 28 (04) : 381 - 387
  • [28] NEW LOW-PRESSURE IMPACTOR - DETERMINATION OF SIZE DISTRIBUTION AEROSOL SULFUR-COMPOUNDS
    HERING, SV
    FLAGAN, RC
    FRIEDLANDER, SK
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1977, 173 (MAR20): : 40 - 40
  • [29] Electrical low pressure impactor
    Keskinen, J., 1600, (23):
  • [30] PERFORMANCE OF A NEW PACKING WITH LOW-PRESSURE DROP
    OZBELGE, HO
    SOMER, TG
    CHEMICAL ENGINEERING JOURNAL AND THE BIOCHEMICAL ENGINEERING JOURNAL, 1990, 45 (01): : 79 - 81