Impact of 6MV photon beam attenuation by carbon fiber couch and immobilization devices in IMRT planning and dose delivery

被引:26
|
作者
Munjal, R. K. [1 ]
Negi, P. S. [1 ]
Babu, A. G. [1 ]
Sinha, S. N. [1 ]
Anand, A. K. [1 ]
Kataria, T. [1 ]
机构
[1] Rajiv Gandhi Canc Inst & Res Ctr, Delhi, India
关键词
Attenuation; base plate; carbon fiber; dosimetry; intensity modulated radiation therapy; treatment couch;
D O I
10.4103/0971-6203.26690
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Multiple fields in IMRT and optimization allow conformal dose to the target and reduced dose to the surroundings and the regions of interest. Thus we can escalate the dose to the target to achieve better tumor control with low morbidity. Orientation of multiple beams can be achieved by i) different gantry angles, ii) rotating patient's couch isocentrically. In doing so, one or more beam may pass through different materials like the treatment couch, immobilization cast fixation plate, head and neck rest or any other supportive device. Our observations for 6MV photon beam on PRIMUS-KXE2 with MED-TEC carbon fiber tabletop and 10 x 10 cm(2) field size reveals that the maximum dose attenuation by the couch was of the order of 2.96% from gantry angle 120-160 degrees. Attenuation due to cast fixation base plate of PMMA alone was of the order of 5.8-10.55% at gantry angle between 0 and 90 degrees. Attenuation due to carbon fiber base plate alone was 3.8-7.98%. Attenuation coefficient of carbon fiber and PMMA was evaluated and was of the order of 0.082 cm(-1) and 0.064 cm(-1) respectively. Most of the TPS are configured for direct beam incidence attenuation correction factors only. Whereas when the beam is obliquely incident on the couch, base plate, headrest and any other immobilization device get attenuated more than the direct beam incidence. The correction factors for oblique incidence beam attenuation are not configured in most of the commercially available treatment planning systems. Therefore, such high variations in dose delivery could lead to under-dosage to the target volume for treatments requiring multiple fields in IMRT and 3D-CRT and need to be corrected for monitor unit calculations.
引用
收藏
页码:67 / 71
页数:5
相关论文
共 30 条
  • [1] The significance of carbon fiber couch top attenuation in a 6MV VMAT beam
    McCormack, S.
    Bee, G.
    Morgan, A. M.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S151 - S151
  • [2] IMRT treatment of prostate cancer using 6MV photon beam
    Huang, J
    Katz, A
    Haas, J
    Urrutia, T
    Lauritano, J
    MacMelville, W
    MEDICAL PHYSICS, 2002, 29 (06) : 1254 - 1254
  • [3] 6MV photon dMLC IMRT: near surface dose with oblique incidence
    Popescu, C.
    Beckham, W.
    Bush, K.
    MEDICAL PHYSICS, 2006, 33 (07) : 2657 - 2657
  • [4] COMPUTER GENERATED SCATTER DOSE DISTRIBUTIONS FOR A 6MV RADIOTHERAPY PHOTON BEAM
    BLOCH, P
    WALLACE, R
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 : 74 - 74
  • [5] Impact of radiation attenuation by a carbon fiber couch on patient dose verification
    Yu, Chun-Yen
    Chou, Wen-Tsae
    Liao, Yi-Jen
    Lee, Jeng-Hung
    Liang, Ji-An
    Hsu, Shih-Ming
    SCIENTIFIC REPORTS, 2017, 7
  • [6] Impact of radiation attenuation by a carbon fiber couch on patient dose verification
    Chun-Yen Yu
    Wen-Tsae Chou
    Yi-Jen Liao
    Jeng-Hung Lee
    Ji-An Liang
    Shih-Ming Hsu
    Scientific Reports, 7
  • [7] Influence of RF Coil Materials On Surface and Buildup Dose From a 6MV Photon Beam
    Ghila, A.
    Fallone, B.
    Rathee, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3321 - 3321
  • [8] The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert
    McCormack, S
    Diffey, J
    Morgan, A
    MEDICAL PHYSICS, 2005, 32 (02) : 483 - 487
  • [9] The effect of gantry angle on megavoltage photon beam attenuation by carbon fiber couch inserts
    McCormack, S
    Diffey, J
    Morgan, A
    MEDICAL PHYSICS, 2004, 31 (06) : 1813 - 1813
  • [10] High photon beam attenuation by the treatment couch and (commercial) immobilisation devices in IMRT fields with "unconventional" angles of incidence
    Vieira, S
    Kaatee, RSJ
    Dirkx, MLP
    Heijmen, BJM
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S37 - S37