AUTOMATIC SIMULTANEOUS DETERMINATION OF CA AND MG IN NATURAL-WATERS WITH NO INTERFERENCE SEPARATION

被引:22
|
作者
RUISANCHEZ, I [1 ]
RIUS, A [1 ]
LARRECHI, MS [1 ]
CALLAO, MP [1 ]
RIUS, FX [1 ]
机构
[1] UNIV ROVIRA & VIRGILI TARRAGONA,DEPT CHEM,PL IMPERIAL TARRACO 1,E-43005 TARRAGONA,SPAIN
关键词
D O I
10.1016/0169-7439(94)00002-6
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this study a methodology is reported for the automatic, simultaneous determination of Ca and Mg in natural waters which is not affected by the presence of species that may interfere with other methods. A flow system based on the sinusoidal injection analysis (SIA) methodology with diode array spectrophotometric detection of the complex formed by both analytes with the Arsenazo III has been used. The technique of multivariate calibration used, PLS1, enables Ca to be determined in the concentration range 45-85 mg l-1 and Mg to be determined in the range 2-70 mg l-1, with no need for the samples to be previously diluted. The accuracy of the method has been tested by comparing the results obtained with the standard atomic absorption spectrometry method. With an analysis speed of 20-30 samples per hour, the RMSE in terms of standard deviation of the original variables was 6.1 for Mg and 6.4 for Ca.
引用
收藏
页码:55 / 63
页数:9
相关论文
共 50 条