A 0.6-TESLA SUPERCONDUCTING MAGNET FOR MAGNETIC-RESONANCE-IMAGING

被引:0
|
作者
ZHANG, CJ
LIN, MH
CHEN, HS
ZHOU, XT
LIU, QN
机构
[1] China Kejian Corporation limited, Shenzhen, SheKou
关键词
2;
D O I
10.1109/20.119956
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The paper introduces the first superconducting magnet system for whole body magnetic resonance imaging built in China. The character and composition of the system was described in detail. Warm bore diameter of the magnet system is 1.08m, the length is 2.55m and total weight is 7000kg. The main magnet is composed of six superconducting coils and contributed 0.6T field. The coils are wound with NbTi wire with 1mm diameter and the rated current is 142.6A. Passive shims are used to correct the central field and achieve the uniformity of 2.5ppm in 30cm DSV. The cryostat system is equipped with LHe and LN2 closed-loop refrigerator which have been developed for MRI by CVI.
引用
收藏
页码:631 / 632
页数:2
相关论文
共 50 条
  • [1] A 4.7 TESLA MAGNET FOR MAGNETIC-RESONANCE-IMAGING AND SPECTROSCOPY
    GREEN, MA
    CAROLAN, J
    IEEE TRANSACTIONS ON MAGNETICS, 1987, 23 (02) : 1299 - 1302
  • [2] Design of the superconducting magnet for 9.4 Tesla whole body magnetic resonance imaging
    Li, Y.
    Wang, Q.
    Dai, Y.
    Ni, Z.
    Zhu, X.
    Li, L.
    Mao, B.
    Chen, S.
    26TH INTERNATIONAL CRYOGENIC ENGINEERING CONFERENCE & INTERNATIONAL CRYOGENIC MATERIALS CONFERENCE 2016, 2017, 171
  • [3] A 60 CM BORE 2.0 TESLA HIGH HOMOGENEITY MAGNET FOR MAGNETIC-RESONANCE-IMAGING
    BOBROV, ES
    PILLSBURY, RD
    PUNCHARD, WFB
    SCHWALL, RE
    SEGAL, HR
    WILLIAMS, JEC
    NEURINGER, LJ
    IEEE TRANSACTIONS ON MAGNETICS, 1987, 23 (02) : 1303 - 1308
  • [4] SUPERCONDUCTING MAGNETS FOR MAGNETIC-RESONANCE-IMAGING
    PITTARD, S
    FEENAN, P
    VENNART, W
    APPLIED SUPERCONDUCTIVITY, 1993, 1 (10-12) : 1827 - 1840
  • [5] 0.6 TESLA MAGNETIC-RESONANCE IMAGING OF THE ORBIT
    EDWARDS, JH
    HYMAN, RA
    VACIRCA, SJ
    BOXER, M
    PACKER, S
    KAUFMAN, IH
    STEIN, HL
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1984, 5 (05) : 681 - 681
  • [6] MAGNETIC-RESONANCE-IMAGING OF INTRACRANIAL NEOPLASMS AT 0.02 TESLA
    KORMANO, M
    RAININKO, R
    KATEVUO, K
    ACTA RADIOLOGICA, 1987, 28 (04) : 369 - 373
  • [7] SUPERCONDUCTING MAGNET WITH SELF-SHIELD FOR WHOLE-BODY MAGNETIC-RESONANCE-IMAGING
    SHIMADA, Y
    MATSUMOTO, T
    MORITSU, K
    TAKECHI, M
    WATANABE, T
    IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 1685 - 1688
  • [8] SUPERCONDUCTING MAGNETS FOR MAGNETIC-RESONANCE-IMAGING APPLICATIONS
    MORROW, GR
    ROSNER, CH
    IEEE TRANSACTIONS ON MAGNETICS, 1987, 23 (02) : 1294 - 1298
  • [9] Superconducting Magnet of Magnetic Resonance Imaging System
    Wang, Zan Ming
    2013 IEEE International Conference on Applied Superconductivity and Electromagnetic Devices (ASEMD), 2013, : 530 - 533
  • [10] TECHNICAL ASPECTS ON MAGNETIC-RESONANCE-IMAGING OF THE SPINE AT 1.5 TESLA
    HOLTAS, SL
    PLEWES, DB
    SIMON, JH
    EKHOLM, S
    KIDO, DK
    UTZ, R
    ACTA RADIOLOGICA, 1987, 28 (04) : 375 - 381