Central corneal thickness and anterior chamber depth measurement by Sirius (R) Scheimpflug tomography and ultrasound

被引:21
|
作者
Jorge, J. [1 ]
Rosado, J. L. [2 ]
Diaz-Rey, J. A. [1 ]
Gonzalez-Meijome, J. M. [1 ]
机构
[1] Univ Minho, Ctr Phys Optometry, Clin & Expt Optometry Res Lab, P-4710057 Braga, Portugal
[2] Opticlin, Lisbon, Portugal
来源
CLINICAL OPHTHALMOLOGY | 2013年 / 7卷
关键词
Scheimpflug corneal tomography; ultrasound biometry; ultrasound pachymetry; limits of agreement;
D O I
10.2147/OPTH.S35121
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: The purpose of this study was to compare the accuracy of the new Sirius (R) Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measurements obtained by ultrasound biometry, respectively. Methods: CCT and ACD was measured in 50 right eyes from 50 healthy subjects using a Sirius Scheimpflug camera, SP100 ultrasound pachymetry, and US800 ultrasound biometry. Results: CCT measured with the Sirius was 546 +/- 39 mu m and 541 +/- 35 mu m with SP100 ultrasound pachymetry (P = 0.003). The difference was statistically significant (mean difference 4.68 +/- 10.5 mu m; limits of agreement -15.8 to 25.20 mu m). ACD measured with the Sirius was 2.96 +/- 0.3 mm compared with 3.36 +/- 0.29 mm using US800 ultrasound biometry (P < 0.001). The difference was statistically significant (mean difference -0.40 +/- 0.16 mm; limits of agreement -0.72 to 0.07 mm). When the ACD values obtained using ultrasound biometry were corrected according to the values for CCT measured by ultrasound, the agreement increased significantly between both technologies for ACD measurements (mean difference 0.15 +/- 0.16 mm; limits of agreement -0.16 to 0.45 mm). Conclusion: CCT and ACD measured by Sirius and ultrasound methods showing good agreement between repeated measurements obtained in the same subjects (repeatability) with either instrument. However, CCT and ACD values, even after correcting ultrasound ACD by subtracting the CCT value obtained with either technology should not be used interchangeably.
引用
收藏
页码:417 / 422
页数:6
相关论文
共 50 条
  • [1] Comparison of Central Corneal Thickness, Thinnest Corneal Thickness, Anterior Chamber Depth, and Simulated Keratometry Using Galilei, Pentacam, and Sirius Devices
    Anayol, Mustafa A.
    Guler, Emre
    Yagci, Ramazan
    Sekeroglu, Mehmet A.
    Yilmazoglu, Meltem
    Tirhis, Hakan
    Kulak, Ali E.
    Yilmazbas, Pelin
    [J]. CORNEA, 2014, 33 (06) : 582 - 586
  • [2] Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth
    Bedei, Andrea
    Appolloni, Ilaria
    Madesani, Antonio
    Pietrelli, Alessia
    Franceschi, Sara
    Barabesi, Lucio
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2012, 22 : S29 - S32
  • [3] Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
    Yili Jin
    Colm McAlinden
    Yong Sun
    Daizong Wen
    Yiran Wang
    Jinjin Yu
    Ke Feng
    Benhao Song
    Qinmei Wang
    Shihao Chen
    Jinhai Huang
    [J]. Eye and Vision, 8
  • [4] Scheimpflug Corneal Tomography Anterior Chamber Depth in Descemet Stripping Endothelial Keratoplasty Recipients
    Mandell, Jordan
    Pajek, Sarah
    Vanner, Elizabeth A.
    Santiago-Leon, Andrea C.
    Zhang, Crystal L.
    Koo, Ellen H.
    [J]. CORNEA, 2024, 43 (09) : 1095 - 1099
  • [5] Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
    Jin, Yili
    McAlinden, Colm
    Sun, Yong
    Wen, Daizong
    Wang, Yiran
    Yu, Jinjin
    Feng, Ke
    Song, Benhao
    Wang, Qinmei
    Chen, Shihao
    Huang, Jinhai
    [J]. EYE AND VISION, 2021, 8 (01)
  • [6] Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness
    Li, Xuepei
    Zhou, Yijing
    Young, Charlotte Aimee
    Chen, Aiming
    Jin, Guangming
    Zheng, Danying
    [J]. ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (14)
  • [7] Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT
    Kuerten, David
    Plange, Niklas
    Koch, Eva Charlotte
    Koutsonas, Antonis
    Walter, Peter
    Fuest, Matthias
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2015, 253 (07) : 1105 - 1109
  • [8] Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT
    David Kuerten
    Niklas Plange
    Eva Charlotte Koch
    Antonis Koutsonas
    Peter Walter
    Matthias Fuest
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2015, 253 : 1105 - 1109
  • [9] Corneal thickness and anterior chamber depth of the normal adult horse as measured by ultrasound biomicroscopy
    Knickelbein, Kelly E.
    Lassaline, Mary E.
    Kim, Soohyun
    Scharbrough, Machal S.
    Thomasy, Sara M.
    [J]. VETERINARY OPHTHALMOLOGY, 2022, 25 : 17 - 24
  • [10] Comparison of Keratometry, Central Corneal Thickness, and Anterior Chamber Depth Results Measured With Nidek-AL Scan Biometry and Sirius Topography Devices
    Bayramoglu, Sadik Etka
    Sayin, Nihat
    Ekinci, Dilbade Yildiz
    Erdogan, Mehmet
    [J]. ISTANBUL MEDICAL JOURNAL, 2018, 19 (02): : 158 - 161